评估 NicAlert 条带的唾液可替宁读数与经自我报告不吸烟但通过一氧化碳但未通过 NicAlert 验证的个体的液相色谱串联质谱检测的准确性。

Assessment of the accuracy of salivary cotinine readings from NicAlert strips against a liquid chromatography tandem mass spectrometry assay in self-reported non-smokers who passed carbon monoxide but failed NicAlert validation.

机构信息

Institute of Global Health, Faculty of Medicine, University of Geneva, Switzerland.

出版信息

Addiction. 2019 Dec;114(12):2252-2256. doi: 10.1111/add.14775. Epub 2019 Aug 26.

Abstract

BACKGROUND AND AIMS

NicAlert produces semi-quantitative assessments of cotinine levels in saliva or urine for verification of smoking abstinence. This study aimed to assess the accuracy of NicAlert readings against a liquid chromatography tandem mass spectrometry assay in smokers who had passed expired-air carbon monoxide (CO) verification but failed NicAlert verification.

DESIGN

Comparison of NicAlert readings against readings from a reference assay using liquid chromatography tandem mass spectrometry.

SETTING

Geneva, Switzerland.

PARTICIPANTS

Self-reported non-smokers (n = 92) who in previous testing had CO of 0-3 parts per million (indicating no recent smoking) and reported not using any nicotine product, but had NicAlert readings ≥ 1 (indicating smoking).

MEASUREMENTS

NicAlert produces readings of 0, 1 and 2+, which are reported by the manufacturer to correspond to saliva cotinine concentrations of 0-10 ng/ml (indicating not smoking), 10-30 ng/ml and 30+ ng/ml, respectively. Liquid chromatography tandem mass spectrometry was used as the reference.

FINDINGS

For 82 participants with a NicAlert reading of 1, only two of the liquid chromatography tandem mass spectrometry values were within the purported range of 10-30 ng/ml; 71 were below 4 ng/ml and half the values were below 0.5 ng/ml. Two of the eight participants with NicAlert readings of 2 had laboratory values within the designated range. Neither of the two participants with NicAlert readings of 3 had a cotinine value within the designated range.

CONCLUSIONS

In people who had passed carbon monoxide verification, NicAlert readings yielded a very high false-positive rate in detecting levels of cotinine indicative of smoking.

摘要

背景和目的

NicAlert 可对半定量评估唾液或尿液中的可铁宁水平,以验证戒烟情况。本研究旨在评估通过呼出一氧化碳(CO)验证但未通过 NicAlert 验证的吸烟者的 NicAlert 读数与液相色谱串联质谱分析的准确性。

设计

将 NicAlert 读数与液相色谱串联质谱参考分析进行比较。

地点

瑞士日内瓦。

参与者

自我报告的非吸烟者(n=92),他们在之前的测试中 CO 为 0-3ppm(表明最近未吸烟),并报告未使用任何尼古丁产品,但 NicAlert 读数≥1(表明吸烟)。

测量

NicAlert 产生 0、1 和 2+读数,制造商报告这些读数分别对应于唾液可铁宁浓度为 0-10ng/ml(表示未吸烟)、10-30ng/ml 和 30+ng/ml。液相色谱串联质谱分析被用作参考。

结果

对于 82 名 NicAlert 读数为 1 的参与者,只有 2 个液相色谱串联质谱值在 10-30ng/ml 的推测范围内;71 个值低于 4ng/ml,一半的值低于 0.5ng/ml。8 名 NicAlert 读数为 2 的参与者中有 2 名的实验室值在指定范围内。NicAlert 读数为 3 的两名参与者均无可铁宁值在指定范围内。

结论

在通过 CO 验证的人群中,NicAlert 读数在检测可铁宁水平以指示吸烟方面产生了非常高的假阳性率。