Mammography Dose Survey Using International Quality Standards.

作者信息

Boujemaa Soumaya, Bosmans Hilde, Bentayeb Farida

机构信息

Faculty of Sciences, University Mohammed V, Rabat, Morocco.

Medical Imaging Research Center, Department of Radiology, University Hospitals Leuven, Leuven, Belgium.

出版信息

J Med Imaging Radiat Sci. 2019 Dec;50(4):529-535. doi: 10.1016/j.jmir.2019.07.003. Epub 2019 Aug 13.

Abstract

Digital mammography is the current standard for breast cancer screening. The absence of any dosimetric data, the quality standards, and the fear of radiation detriment, sometimes, hampers the smooth introduction of this technology and the launch of breast cancer screening programmes. As the breast cancer screening programmes are in development in Morocco, quality standards have been set in this study, so that any new breast cancer screening service could get started with their first analysis. The purpose of this study was to report the first Moroccan breast dose and to test quality standards compared with European guidelines. A dosimetric study is conducted by calculating mean glandular dose (MGD) for patients' breasts and polymethyl methacrylate (PMMA) for each thickness from recorded radiographic factors and X-ray tube (output and half-value layer) measurements using Dance's model for a digital mammography system. This is carried out to determine the correlation between phantom and patient measurements. The mean MGD for patient data was 1.02 ± 0.5 mGy and the compressed breast thickness was 55 ± 14 mm. For compressed breast thickness range of 50-60 mm, the MGD was 0.94 ± 0.3 mGy and the dose calculated with PMMA phantom for the same range thickness was 1.33 mGy. The results of MGD as a function of PMMA thickness is lower than the achievable limit curve proposed in the European guidelines. The average doses, which depend on the technical parameters of the mammography equipment, are in line with values obtained in European guidelines. In addition, an investigation is needed throughout the Moroccan territory, in the sense of optimization of radiological practices and techniques in mammographic centres aligning with the international recommendations.

摘要

数字化乳腺摄影是目前乳腺癌筛查的标准方法。有时,由于缺乏任何剂量数据、质量标准以及对辐射危害的担忧,会阻碍这项技术的顺利引入和乳腺癌筛查项目的开展。鉴于摩洛哥的乳腺癌筛查项目正在发展中,本研究制定了质量标准,以便任何新的乳腺癌筛查服务在首次分析时就能启动。本研究的目的是报告摩洛哥的首次乳腺剂量,并与欧洲指南相比测试质量标准。通过使用数字乳腺摄影系统的丹斯模型,根据记录的射线照相因素和X射线管(输出和半价层)测量值,计算患者乳房的平均腺体剂量(MGD)和每种厚度的聚甲基丙烯酸甲酯(PMMA)的剂量,进行剂量学研究。这样做是为了确定体模测量和患者测量之间的相关性。患者数据的平均MGD为1.02±0.5毫戈瑞,压缩乳房厚度为55±14毫米。对于50 - 6   毫米的压缩乳房厚度范围,MGD为0.94±0.3毫戈瑞,相同厚度范围的PMMA体模计算剂量为1.33毫戈瑞。MGD作为PMMA厚度函数的结果低于欧洲指南中提出的可达到的极限曲线。平均剂量取决于乳腺摄影设备的技术参数,与欧洲指南中获得的值一致。此外,有必要在摩洛哥全国范围内进行调查,以优化乳腺摄影中心的放射学实践和技术,使其符合国际建议。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验