As the total volume of mammograms in Dubai is increasing consistently, it is crucial to focus on the process of dose optimization by determining dose reference levels for such sensitive radiographic examinations as mammography. This work aimed to determine local diagnostic reference levels (DRLs) for mammography procedures in Dubai at different ranges of breast thickness. A total of 2599 anonymized mammograms were randomly retrieved from a central dose survey database. Mammographic cases for screening women aged from 40 to 69 years were included, while cases of breast implants and breast thickness outside the range of 20-100 mm were excluded. Mean, median, and 75 percentiles were obtained for the mean glandular dose (MGD) distribution of each mammography projection for all compressed breast thickness (CBT) ranges. The local DRLs for mammography in Dubai were found to be between 0.80 mGy and 0.82 mGy for the craniocaudal (CC) projection and between 0.89 mGy and 0.971.8 mGy for the mediolateral oblique (MLO) projection. Local DRLs were proposed according to different breast thicknesses, starting from 20 to 100 mm. All groups of CBT showed a slight difference in MGD values, with higher values in MLO views rather than CC views. The local DRLs in this study were lower than some other Middle Eastern countries and lower than the standard reference levels reported by the International Atomic Energy Agency (IAEA) at 3 mGy/view.
由于迪拜乳房X光检查的总量持续增加,通过确定乳房X光摄影等敏感放射检查的剂量参考水平来关注剂量优化过程至关重要。这项工作旨在确定迪拜不同乳房厚度范围内乳房X光检查程序的本地诊断参考水平(DRL)。从中央剂量调查数据库中随机检索了总共2599份匿名乳房X光片。纳入了40至69岁筛查女性的乳房X光检查病例,同时排除了乳房植入物病例和乳房厚度在20 - 100毫米范围之外的病例。针对所有压缩乳房厚度(CBT)范围,获得了每个乳房X光摄影投影的平均腺体剂量(MGD)分布的均值、中位数和第75百分位数。发现迪拜乳房X光摄影的本地DRL在头尾位(CC)投影时为0.80毫戈瑞至0.82毫戈瑞之间,在内外侧斜位(MLO)投影时为0.89毫戈瑞至0.971.8毫戈瑞之间。根据不同的乳房厚度(从20毫米到100毫米)提出了本地DRL。所有CBT组的MGD值略有差异,MLO视图中的值高于CC视图。本研究中的本地DRL低于其他一些中东国家,也低于国际原子能机构(IAEA)报告的3毫戈瑞/视图的标准参考水平。