J Strength Cond Res. 2021 Oct 1;35(10):2894-2901. doi: 10.1519/JSC.0000000000003211.
Lunn, WR and Axtell, RS. Validity and reliability of the Lode Excalibur Sport cycle ergometer for the Wingate Anaerobic Test. J Strength Cond Res 35(10): 2894-2901, 2021-Although multiple testing devices advertise Wingate Anaerobic Test (WAnT), capability, reliability, and validity data are sparse. The purpose was to determine whether the Lode Excalibur Sport cycle ergometer is a reliable and valid instrument to conduct the 30-second WAnT when compared with the Monark 894e Peak Bike ergometer. Recreationally active men (n = 49; 20.6 ± 2.5 years; 1.75 ± 0.07 m; and 79.1 ± 9.8 kg) completed four 30-second WAnTs: 2 each on the Lode and the Monark ergometers for peak power (PP), mean power (MP), minimum power (MinP), fatigue index (FI), and peak cadence (vmax) measurement. Interday and interunit correlation, multivariate analysis of variance, regression, and Bland-Altman analysis determined reliability and validity. Cohen's d was used to determine effect size. Relative reliability (intraclass correlation coefficient) with 95% confidence interval for Monark and Lode was very high for PP, MP, MinP, and vmax and high for FI (r > 0.83; coefficient of variation ≤ 27.0%; p < 0.01). Interunit correlation was strong for PP, MP, MinP, and vmax (r > 0.75; p < 0.001) and moderate for FI (p = 0.001). Lode PP and FI values were significantly less (p < 0.001; d > 1.18) and MinP, MP, and vmax significantly greater (p ≤ 0.001; d > 0.51) than Monark. Proportional bias was demonstrated for all variables (p < 0.04; d > 2.68) except vmax. The Lode ergometer reliably provides WAnT outcomes and correlates well to the Monark ergometer. However, differences in power values and proportional bias between differently braked instruments prevent use of the Lode ergometer for comparison of WAnT data with normative data generated by the Monark ergometer.
伦特,WR 和阿克斯特尔,RS。洛德勇士运动脚踏车测功计用于瓦格纳无氧测试的有效性和可靠性。J 强度条件研究 35(10):2894-2901,2021-尽管有多种测试设备宣传瓦格纳无氧测试(WAnT),但能力、可靠性和有效性数据很少。目的是确定洛德勇士运动脚踏车测功计是否可以与 Monark 894e 峰值自行车测功计一起可靠且有效地进行 30 秒的 WAnT。有休闲活动的男性(n = 49;20.6 ± 2.5 岁;1.75 ± 0.07 m;79.1 ± 9.8 kg)完成了四次 30 秒的 WAnT:洛德和 Monark 测功计各两次,用于测量峰值功率(PP)、平均功率(MP)、最小功率(MinP)、疲劳指数(FI)和峰值踏频(vmax)。日内和单元间相关性、多变量方差分析、回归和 Bland-Altman 分析确定了可靠性和有效性。使用 Cohen's d 确定效应大小。Monark 和 Lode 的可靠性(组内相关系数)的 95%置信区间非常高,PP、MP、MinP 和 vmax 高,FI 高(r > 0.83;变异系数≤27.0%;p < 0.01)。单元间相关性对 PP、MP、MinP 和 vmax 很强(r > 0.75;p < 0.001),对 FI 适中(p = 0.001)。Lode 的 PP 和 FI 值显著低于(p < 0.001;d > 1.18),MinP、MP 和 vmax 值显著高于(p ≤ 0.001;d > 0.51)Monark。所有变量都表现出比例偏差(p < 0.04;d > 2.68),除了 vmax。Lode 测功计可靠地提供 WAnT 结果,与 Monark 测功计相关性良好。然而,不同制动仪器之间的功率值和比例偏差的差异阻止了使用 Lode 测功计将 WAnT 数据与 Monark 测功计生成的规范数据进行比较。