J Clin Monit Comput. 2020 Aug;34(4):649-654. doi: 10.1007/s10877-019-00375-z. Epub 2019 Aug 27.
Pulse wave analysis enables stroke volume to be estimated from an arterial blood pressure waveform. Multi-beat analysis is a novel pulse wave analysis method. We aimed to investigate cardiac output (CO) estimations using multi-beat analysis of the radial arterial blood pressure waveform in patients undergoing off-pump coronary artery bypass surgery (OPCAB) using intermittent pulmonary artery thermodilution (PATD) as the reference method. This was a prospective clinical method comparison study. In 58 patients, we measured CO using PATD (PATD-CO; reference method) and simultaneously recorded the radial arterial blood pressure waveform that we used for off-line estimation of CO based on multi-beat analysis (MBA-CO; test method) using the Argos CO monitor (Retia Medical; Valhalla, NY, USA). The final analysis was performed using 572 paired CO measurements. We performed Bland-Altman analysis accounting for multiple observations per patient. To describe the ability of the test method to track changes in CO over time we computed four-quadrant plots using a central exclusion zone of 15% and calculated the concordance rate. Mean PATD-CO was 4.13 ± 1.26 L/min and mean MBA-CO was 4.31 ± 1.25 L/min. The mean of the differences between PATD-CO and MBA-CO was - 0.20 L/min with a standard deviation of ± 1.14 L/min and 95% limits of agreement of - 2.48 to + 2.08 L/min. The concordance rate for CO changes between PATD-CO and MBA-CO was 89%. CO estimations using multi-beat analysis (Argos monitor) show reasonable agreement and trending ability compared with PATD-CO as the reference method in adult patients during OPCAB.
脉搏波分析可根据动脉血压波形估计每搏输出量。多节拍分析是一种新的脉搏波分析方法。我们旨在研究使用间歇肺动脉热稀释法(PATD)作为参考方法的非体外循环冠状动脉旁路移植术(OPCAB)患者桡动脉血压波形的多节拍分析来估计心输出量(CO)。这是一项前瞻性临床方法比较研究。在 58 例患者中,我们使用 PATD(PATD-CO;参考方法)测量 CO,并同时记录桡动脉血压波形,我们使用 Argos CO 监测仪(Retia Medical;纽约州瓦哈拉)对其进行离线基于多节拍分析的 CO 估计(MBA-CO;测试方法)。最终分析使用了 572 对配对的 CO 测量值。我们进行了 Bland-Altman 分析,考虑了每位患者的多次观察。为了描述测试方法随时间跟踪 CO 变化的能力,我们使用中央排除区为 15%的四象限图进行了计算,并计算了一致性率。PATD-CO 的平均值为 4.13±1.26 L/min,MBA-CO 的平均值为 4.31±1.25 L/min。PATD-CO 和 MBA-CO 之间差值的平均值为 -0.20 L/min,标准差为±1.14 L/min,95%一致性界限为 -2.48 至 +2.08 L/min。PATD-CO 和 MBA-CO 之间 CO 变化的一致性率为 89%。在 OPCAB 期间,与作为参考方法的 PATD-CO 相比,使用多节拍分析(Argos 监测仪)进行 CO 估计显示出合理的一致性和趋势能力。