Faculty of Applied Medical Sciences, Jordan University of Science and Technology, Irbid, Jordan.
Department of Diagnostic Radiology, American University of Beirut Medical Center, Beirut, Lebanon.
J Appl Clin Med Phys. 2019 Oct;20(10):181-186. doi: 10.1002/acm2.12711. Epub 2019 Aug 30.
The aim of this study was to determine diagnostic reference levels (DRLs) for cardiac computed tomography (CCT) in Jordan. Volume computed tomography dose index (CTDI ) and dose-length product (DLP) were collected from 228 CCTs performed at seven Jordanian hospitals specialized in cardiac CT. DRLs for cardiac CT were defined at the 75th percentile of CTDI and DLP. CTDI and DLP were collected from 30 successive cardiac CT in each center except for one center (18 scans). The 75th percentile of the CTDI and the DLP of the centers calculated from mixed retrospective and prospective gated modes were 47.74 milligray (mGy) and 1035 mGy/cm, respectively. This study demonstrated wide dose variations among the surveyed hospitals for cardiac CT scans; there was a 5.1-fold difference between the highest and lowest median DLP with a range of 223.2-1146.7 mGy/cm. Differences were associated with variations in the mAs and kVp. This study confirmed large variability in CTDI and DLP for cardiac CT scans; variation was associated with acquisition protocols and highlights the need for dose optimization. DRLs are proposed for CCT; there remains substantial potential for optimization of cardiac CT examinations for adults in Jordan.
本研究旨在确定约旦心脏 CT(CCT)的诊断参考水平(DRL)。从约旦 7 家专门从事心脏 CT 的医院进行的 228 次 CCT 中收集了容积 CT 剂量指数(CTDI)和剂量长度乘积(DLP)。心脏 CT 的 DRL 定义为 CTDI 和 DLP 的第 75 百分位数。除了一个中心(18 次扫描)外,每个中心都从混合回顾性和前瞻性门控模式中收集了 30 次连续心脏 CT 的 CTDI 和 DLP。来自混合回顾性和前瞻性门控模式的中心的 CTDI 和 DLP 的第 75 百分位数分别为 47.74 毫戈瑞(mGy)和 1035 mGy/cm。本研究表明,接受调查的医院之间心脏 CT 扫描的剂量差异很大;最高和最低中位数 DLP 之间存在 5.1 倍的差异,范围为 223.2-1146.7 mGy/cm。差异与 mAs 和 kVp 的变化有关。本研究证实了心脏 CT 扫描的 CTDI 和 DLP 存在很大的可变性;变异性与采集协议有关,突出了剂量优化的必要性。为 CCT 提出了 DRL;约旦成年人的心脏 CT 检查仍有很大的优化潜力。