Abstract

BACKGROUND

Proflexendo file (Nexen, Houston, TX, USA) is a recently developed, novel nickel-titanium rotary file. However, several physical properties of the Proflexendo system remain to be investigated.

OBJECTIVES

The aim of this study was to determine the effect of sterilization on cyclic resistance of Proflexendo files.

MATERIAL AND METHODS

A total of 120 unused Proflexendo rotary files (40 each of file sizes 40/0.04, 30/0.04 and 20/0.06) were used in this study. Each set of files was subdivided into four groups of ten files. Group 1 did not undergo any sterilization (control), Group 2 underwent two rounds (2×) of sterilization, Group 3 five rounds (5×) and Group 4 ten rounds (10×). Cyclic fatigue resistance was tested using an artificial canal with a 5-mm radius curve.

RESULTS

The nonsterilized size 30/0.04 files had the highest cycles to failure, followed by the nonsterilized size 40/0.04 and 20/0.06 files ( < 0.001). With increasing rounds of sterilization, cycles to failure reduced for sizes 40 (2×, 5× and 10×) and 30 (2×) files compared with nonsterilized sizes 40 ( < 0.05) and 30 files ( < 0.001), respectively.

CONCLUSION

Under the conditions of the current study, the results provide preliminary evidence that autoclave sterilization of Proflexendo rotary files reduced their cyclic fatigue resistance, except for size 20 (2× and 5×) and 30 (10×) files, in which resistance increased. Single use of this file is recommended to reduce the risk of separation.

摘要

背景

Proflexendo锉(美国得克萨斯州休斯顿市的Nexen公司生产)是一种最近研发的新型镍钛旋转锉。然而,Proflexendo系统的一些物理特性仍有待研究。

目的

本研究旨在确定灭菌对Proflexendo锉循环抗力的影响。

材料与方法

本研究共使用了120根未使用过的Proflexendo旋转锉(锉尺寸40/0.04、30/0.04和20/0.06各40根)。每组锉再细分为四组,每组十根。第1组未进行任何灭菌(对照),第2组进行两轮(2×)灭菌,第3组进行五轮(5×)灭菌,第4组进行十轮(10×)灭菌。使用半径为5毫米的人工根管测试循环疲劳抗力。

结果

未灭菌的30/0.04尺寸锉的失效循环次数最高,其次是未灭菌的40/0.04和20/0.06尺寸锉(<0.001)。随着灭菌轮数的增加,与未灭菌的40尺寸锉(<0.05)和30尺寸锉(<0.001)相比,40(2×、5×和10×)和30(2×)尺寸锉的失效循环次数分别减少。

结论

在本研究条件下,结果提供了初步证据,即Proflexendo旋转锉的高压灭菌会降低其循环疲劳抗力,但20(2×和5×)和30(10×)尺寸锉除外,这两种尺寸的锉抗力增加。建议一次性使用该锉以降低分离风险。