Experimental evaluation on the influence of autoclave sterilization on the cyclic fatigue of new nickel-titanium rotary instruments.

机构信息

Department of Endodontics I, Sapienza University of Rome, Rome, Italy.

出版信息

J Endod. 2012 Feb;38(2):222-5. doi: 10.1016/j.joen.2011.10.017. Epub 2011 Nov 21.

Abstract

INTRODUCTION

The purpose of this study was to evaluate the effect of autoclave sterilization on cyclic fatigue resistance of rotary endodontic instruments made of traditional and new nickel-titanium (NiTi) alloys.

METHODS

Four NiTi rotary endodontic instruments of the same size (tip diameter 0.40 mm and constant .04 taper) were selected: K3, Mtwo, Vortex, and K3 XF prototypes. Each group was then divided into 2 subgroups, unsterilized instruments and sterilized instruments. The sterilized instruments were subjected to 10 cycles of autoclave sterilization. Twelve files from each different subgroup were tested for cyclic fatigue resistance. Means and standard deviations of number of cycles to failure (NCF) and fragment length of the fractured tip were calculated for each group, and data were statistically analyzed (P < .05).

RESULTS

Comparing the results between unsterilized and sterilized instruments for each type of file, differences were statistically significant (P < .05) only between sterilized and unsterilized K3XF files (762 versus 651 NCF). The other instruments did not show significant differences (P > .05) in the mean NCF as a result of sterilization cycles (K3, 424 versus 439 NCF; Mtwo, 409 versus 419 NCF; Vortex, 454 versus 480 NCF). Comparing the results among the different groups, K3 XF (either sterilized or not) showed a mean NCF significantly higher than all other files (P < .05).

CONCLUSIONS

Repeated cycles of autoclave sterilization do not seem to influence the mechanical properties of NiTi endodontic instruments except for the K3 XF prototypes of rotary instruments that demonstrated a significant increase of cyclic fatigue resistance.

摘要

简介

本研究旨在评估高压蒸汽灭菌对传统和新型镍钛(NiTi)合金旋转根管器械循环疲劳性能的影响。

方法

选择了四种相同尺寸(尖端直径 0.40 毫米,恒定.04 锥度)的 NiTi 旋转根管器械:K3、Mtwo、Vortex 和 K3 XF 原型。每组再分为 2 个亚组,即未经消毒的器械和消毒的器械。消毒器械接受 10 次高压蒸汽灭菌循环。对每个不同亚组的 12 个器械进行循环疲劳性能测试。计算每组失效循环数(NCF)和断裂尖端碎片长度的平均值和标准差,并进行统计学分析(P <.05)。

结果

比较每种器械未经消毒和消毒后的结果,仅消毒和未消毒的 K3XF 器械之间存在统计学差异(P <.05)(762 与 651 个 NCF)。其他器械由于消毒循环,NCF 的平均值没有显著差异(P >.05)(K3,424 与 439 NCF;Mtwo,409 与 419 NCF;Vortex,454 与 480 NCF)。比较不同组之间的结果,K3 XF(无论是消毒还是未消毒)的平均 NCF 均显著高于所有其他器械(P <.05)。

结论

高压蒸汽灭菌的重复循环似乎不会影响镍钛根管器械的机械性能,除了旋转器械的 K3 XF 原型,其循环疲劳阻力显著增加。