Al-Amidi Arkan H, Al-Gharrawi Hikmet Abdul-Rahim
Alkarkh Health Directorate Baghdad, Ministry of Health, Baghdad, Iraq.
Department of Conservative Dentistry, College of Dentistry, Mustansiriyah University, Baghdad, Iraq.
J Conserv Dent. 2023 Jan-Feb;26(1):26-30. doi: 10.4103/jcd.jcd_611_21. Epub 2022 Aug 17.
Rotary nickel-titanium files have become the most commonly used engine-powered endodontic instruments due to their improvement in fatigue resistance and flexibility. The autoclaving is the most commonly used method for sterilization. However, the risk of instrument fracture, mainly due to cyclic fatigue during rotation within a curved canal, is still a matter of critical concern.
This study aimed to measure and compares the cyclic fatigue resistance of EdgeFile X7, 2Shape, and F-One files in a single curved simulated canal before and after autoclave sterilization.
Twenty-four files each of EdgeFile X7, 2Shape, and F-One files were used in this study. They were grouped as group A: EdgeFile X7, group B: 2Shape, and group C: F-One files. Then, each group was subdivided into two, sterilized and nonsterilized, subgroups. The files were tested using a custom-made artificial canal. The number of cycles to fracture (NCF) was calculated.
One-way analysis of variance, post hoc Tukey's test, and independent -test were used.
In nonsterilized instruments, the NCF of the EdgeFile X7 group was significantly higher than that in the other groups ( < 0.05). In sterilized instruments, there was a nonsignificant difference in NCF between EdgeFile X7 and F-One files ( > 0.05). The NCF of 2Shape was significantly lower than other files ( ≤ 0.05) in sterilized and nonsterilized groups. The autoclaving significantly decreased ( ≤ 0.05) the cyclic fatigue resistance of the tested files.
EdgeFile X7 was the most fatigue resistant. Autoclaving reduced the cyclic fatigue resistance of the tested files.
由于旋转镍钛锉在抗疲劳性和柔韧性方面有所改进,已成为最常用的电动牙髓治疗器械。高压灭菌是最常用的灭菌方法。然而,器械折断的风险,主要是由于在弯曲根管内旋转时的循环疲劳,仍然是一个至关重要的问题。
本研究旨在测量并比较EdgeFile X7、2Shape和F-One锉在模拟单根弯曲根管中经高压灭菌前后的循环疲劳抗力。
本研究使用了EdgeFile X7、2Shape和F-One锉各24根。它们被分为A组:EdgeFile X7;B组:2Shape;C组:F-One锉。然后,每组再细分为灭菌和未灭菌两个亚组。使用定制的人工根管对锉进行测试。计算折断前的循环次数(NCF)。
采用单因素方差分析、事后Tukey检验和独立检验。
在未灭菌的器械中,EdgeFile X7组的NCF显著高于其他组(P<0.05)。在已灭菌的器械中,EdgeFile X7和F-One锉之间的NCF无显著差异(P>0.05)。在灭菌和未灭菌组中,2Shape的NCF均显著低于其他锉(P≤0.05)。高压灭菌显著降低了受试锉的循环疲劳抗力(P≤0.05)。
EdgeFile X7的抗疲劳性最强。高压灭菌降低了受试锉的循环疲劳抗力。