Huang Jian-Feng, Zhang Wei, Li Yan, Wang Ji-Guang
Centre for Epidemiological Studies and Clinical Trials, The Shanghai Institute of Hypertension, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, China.
Blood Press Monit. 2019 Dec;24(6):315-318. doi: 10.1097/MBP.0000000000000405.
To evaluate the accuracy of the YuWell YE680B oscillometric upper-arm blood pressure monitor for clinic and home blood pressure measurements in adults according to the American National Standards Institute/the Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard.
Eighty-five subjects (male 46 and female 39) with a mean age of 44.6 years were recruited to fulfill the age, sex, blood pressure, and cuff distribution criteria of the ANSI/AAMI/ISO standard in the general population using the same arm sequential blood pressure measurement method. Two differently-sized cuffs were used for arm circumferences 22-32 cm (medium) and 22-45 cm (large), respectively, for the test device, and for arm circumferences ≤32 cm (standard) and >32 cm (large), respectively, for the mercury sphygmomanometer.
A total of 255 comparison pairs were obtained for analysis. For validation criterion 1, the mean ± SD of the differences between the test device and reference systolic/diastolic blood pressure readings was -2.2 ± 7.2/-3.9 ± 5.2 mmHg. For validation criterion 2, the SD of the averaged systolic/diastolic blood pressure differences between the test device and reference blood pressure per subject was 5.70/4.67 mmHg.
The automated upper-arm blood pressure monitor YuWell YE680B has passed the requirements of the ANSI/AAMI/ISO81060-2:2013 standard, and hence can be recommended for blood pressure measurement in the general population.
根据美国国家标准学会/医疗仪器促进协会/国际标准化组织(ANSI/AAMI/ISO)81060-2:2013标准,评估鱼跃YE680B振荡式上臂血压计在成人临床和家庭血压测量中的准确性。
招募85名受试者(男性46名,女性39名),平均年龄44.6岁,采用同臂序贯血压测量法,使其符合一般人群中ANSI/AAMI/ISO标准的年龄、性别、血压和袖带分布标准。测试设备分别使用两种不同尺寸的袖带,臂围22 - 32 cm(中号)和22 - 45 cm(大号),水银血压计分别使用臂围≤32 cm(标准)和>32 cm(大号)的袖带。
共获得255对比较数据用于分析。对于验证标准1,测试设备与参考收缩压/舒张压读数之间差异的均值±标准差为-2.2±7.2/-3.9±5.2 mmHg。对于验证标准2,每位受试者测试设备与参考血压之间平均收缩压/舒张压差异的标准差为5.70/4.67 mmHg。
鱼跃自动上臂血压计YE680B已通过ANSI/AAMI/ISO81060-2:2013标准的要求,因此可推荐用于一般人群的血压测量。