Department of Cardiology, Biwako Central Hospital, Otsu City, Shiga.
Department of Technology Development, Omron Healthcare Co, Ltd., Mukou City, Kyoto, Japan.
Blood Press Monit. 2021 Apr 1;26(2):156-159. doi: 10.1097/MBP.0000000000000520.
The present study aimed to evaluate the accuracy of the Omron HEM-7600T automated oscillometric upper-arm blood pressure (BP) monitor according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 protocol.
We recruited individuals until we obtained 86 eligible participants whose BP levels and ranges of arm circumference met the BP distribution requirements specified by the ANSI/AAMI/ISO. In each individual, we sequentially measured systolic and diastolic blood pressures (DBPs) using a mercury sphygmomanometer (two observers) and the HEM-7600T device (one supervisor) in the left upper arm. Data analysis was performed according to the ANSI/AAMI/ISO guidelines.
A total of 101 individuals were recruited, and 86 were finally included in the analysis. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP was -0.6 ± 6.6/1.0 ± 6.5 mmHg (systolic/diastolic). The data were in accordance with criterion 1 of the ANSI/AAMI/ISO 81060-2:2013 standard requirements (≤ 5 ± ≤ 8 mmHg). In addition, the mean device-observer difference in the 86 participants was -0.6 ± 5.65 mmHg for systolic blood pressure and 1.0 ± 5.96 mmHg for DBP, fulfilling criterion 2 with an SD of ≤6.91 mmHg for systolic blood pressure and ≤6.87 mmHg for DBP.
The Omron HEM-7600T BP monitor fulfilled the requirements of the ANSI/AAMI/ISO validation protocol in adults and can be recommended for home BP measurements.
本研究旨在根据美国国家标准协会/医疗器械促进协会/国际标准化组织(ANSI/AAMI/ISO)81060-2:2013 协议评估欧姆龙 HEM-7600T 自动示波法上臂血压(BP)监测仪的准确性。
我们招募了符合 ANSI/AAMI/ISO 规定的 BP 水平和臂围范围要求的 86 名合格参与者,然后对其进行研究。在每位参与者中,我们使用水银血压计(两名观察者)和 HEM-7600T 设备(一名主管)依次测量左上臂的收缩压和舒张压(DBP)。数据分析按照 ANSI/AAMI/ISO 指南进行。
共招募了 101 人,最终有 86 人纳入分析。对于验证标准 1,测试设备与参考 BP 之间的差异平均值±SD 为-0.6±6.6/1.0±6.5mmHg(收缩压/舒张压)。数据符合 ANSI/AAMI/ISO 81060-2:2013 标准要求的标准 1(≤5±≤8mmHg)。此外,86 名参与者中设备与观察者之间的平均差异为收缩压-0.6±5.65mmHg,舒张压 1.0±5.96mmHg,满足收缩压 SD≤6.91mmHg、舒张压 SD≤6.87mmHg 的标准 2。
欧姆龙 HEM-7600T BP 监测仪符合成人 ANSI/AAMI/ISO 验证方案的要求,可推荐用于家庭 BP 测量。