McGough Ellen L, Hsu Lin-Ya, Thompson Hilaire, Teri Linda
Department of Rehabilitation Medicine, University of Washington.
School of Nursing, University of Washington.
Phys Occup Ther Geriatr. 2018;36(4):399-410. doi: 10.1080/02703181.2018.1556231. Epub 2019 Jan 14.
To examine concurrent validity of inertial sensor (APDM ISway) versus force plate center of pressure (COP) measures of postural sway in cognitively impaired older adults.
Participants, mean age 85.6 (SD 4.8), were tested in 4 static standing conditions: (1) eyes open/normal base, (2) eyes open/narrow base, (3) eyes closed/normal base, and (4) eyes closed/narrow base. ISway and COP measures were collected.
Strong correlations between ISway trunk sway smoothness [ISway JERK, and COP path length (r = 0.67-0.85) and COP mean velocity (r = 0.77-0.87); also ISway total sway acceleration path length/trail duration [ISway PATH, (m/s)] and COP path length (r = 0.77-0.87) and COP mean velocity (r = 0.77-0.91). Increased sway was detected in narrow versus normal base and eyes closed versus open conditions ( = .001).
APDM ISway demonstrated concurrent validity to force-plate COP and changes in postural sway were detected between conditions.