测量静态站立任务中姿势摇摆测量的可靠性:年龄的影响。

Measuring the Reliability of Postural Sway Measurements for a Static Standing Task: The Effect of Age.

作者信息

Lo Pei-Yi, Su Bo-Lin, You Yu-Lin, Yen Chen-Wen, Wang Shih-Ting, Guo Lan-Yuen

机构信息

Department of Sports Medicine, College of Medicine, Kaohsiung Medical University, Kaohsiung, Taiwan.

Department of Mechanical and Electro-mechanical Engineering, National Sun Yat-Sen University, Kaohsiung, Taiwan.

出版信息

Front Physiol. 2022 May 13;13:850707. doi: 10.3389/fphys.2022.850707. eCollection 2022.

Abstract

A force plate is used to determine the ability to balance ability. However, only some medical centers or laboratories are equipped with force plates because they are costly so a low-cost force plate is required for home care or health care institutes. Few studies compare the reliability of postural sway measurements in terms of age. This study proposes a low-cost force plate to select reliable parameters to evaluate postural sway. To determine the intra-rater reliability of a novel force plate and the effect of age difference on the intra-rater test-retest reliability for the center of pressure (COP). Forty participants were enrolled for this study: 20 youths and 20 older adults. Participants stood on a custom-made and low-cost force plate with eyes opened and eyes closed to measure COP-related parameters. The within-day test-retest reliability was measured at two sessions on the same day and the between-days reliability was measured on two different days. The COP-related parameters include the average velocity of COP, the average velocity in the antero-posterior and medio-lateral directions, the mean distance of COP and the mean distance in the antero-posterior and medio-lateral directions. An intra-class correlation coefficient test with one-way random model was performed to determine the reliability of different variables within-days and between-days. The results were presented in single measurement of intraclass correlation coefficient (ICC), the standard error of measurements, and the minimal detectable changes of each COP-related parameters. The novel low-cost force plate demonstrates excellent reliability in terms of the COP velocity related parameters for within- and between-day measurements. The ICC of COP distance related parameters were good to excellent reliability for between-day measurements (range: 0.43-0.84). Older adults demonstrated excellent reliability in terms of the mean distance for antero-posterior and the results were better than those for younger participants for the eyes-opened and eyes-closed conditions. The reliability in terms of the mean distance for medio-lateral was poor to good for older adults (range: 0.38-0.55), and excellent for younger participants. The novel and low-cost force plate reliably measured balance and age affects the reliability of different COP variables, so the results of this study were pertinent to the selection of COP measures.

摘要

测力板用于确定平衡能力。然而,由于成本高昂,只有一些医疗中心或实验室配备了测力板,因此家庭护理或医疗保健机构需要低成本的测力板。很少有研究比较姿势摆动测量在年龄方面的可靠性。本研究提出了一种低成本的测力板,以选择可靠的参数来评估姿势摆动。 为了确定新型测力板的评分者内信度以及年龄差异对压力中心(COP)评分者内重测信度的影响。本研究招募了40名参与者:20名年轻人和20名老年人。参与者站在定制的低成本测力板上,睁眼和闭眼以测量与COP相关的参数。在同一天的两个时段测量日内重测信度,在两个不同的日期测量日间信度。与COP相关的参数包括COP的平均速度、前后方向和内外侧方向的平均速度、COP的平均距离以及前后方向和内外侧方向的平均距离。采用单向随机模型的组内相关系数检验来确定不同变量在日内和日间的信度。结果以组内相关系数(ICC)的单次测量、测量的标准误差以及每个与COP相关参数的最小可检测变化来呈现。 新型低成本测力板在日内和日间测量的COP速度相关参数方面表现出出色的信度。COP距离相关参数的ICC在日间测量中具有良好到出色的信度(范围:0.43 - 0.84)。老年人在前后平均距离方面表现出出色的信度,在睁眼和闭眼条件下,结果优于年轻参与者。老年人在内外侧平均距离方面的信度较差到良好(范围:0.38 - 0.55),而年轻参与者则表现出色。 新型低成本测力板可靠地测量了平衡,年龄影响不同COP变量的信度,因此本研究结果与COP测量的选择相关。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/deff/9138610/91ec20f939eb/fphys-13-850707-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验