Laboratory of Microbiology, Mohammed VI University Hospital, Faculty of Medicine and Pharmacy (University Mohammed the First), Oujda, Morocco.
Laboratory of Microbiology, Mohammed VI University Hospital, Faculty of Medicine and Pharmacy (University Mohammed the First), Oujda, Morocco.
Anal Biochem. 2020 Jan 15;589:113501. doi: 10.1016/j.ab.2019.113501. Epub 2019 Nov 5.
We evaluated the body fluid module on Sysmex UF-1000i (UF-1000i-BF) for analysis of white blood cell (WBC) and red blood cell (RBC) in cerebrospinal fluid. We collected 93 cerebrospinal fluid samples and compared the results of the UF-1000i-BF mode with the Fast-Read 102 disposable counting cell. Results shows a good correlation between the UF-1000i and the microscopic examination. The concordance percentage is 99.06% for white blood cells and 85.18% for red blood cells. The UF-1000i-BF mode offers rapid and reliable total WBC and RBC counts for initial screening of cerebrospinal fluid, and can improve the workflow in a routine laboratory.
我们评估了 Sysmex UF-1000i(UF-1000i-BF)上的体液模块,以分析脑脊液中的白细胞(WBC)和红细胞(RBC)。我们收集了 93 份脑脊液样本,并将 UF-1000i-BF 模式的结果与 Fast-Read 102 一次性计数池进行了比较。结果表明 UF-1000i 与显微镜检查之间具有良好的相关性。对于白细胞,一致性百分比为 99.06%,对于红细胞,一致性百分比为 85.18%。UF-1000i-BF 模式可快速可靠地进行总白细胞和红细胞计数,用于初步筛选脑脊液,并且可以改善常规实验室的工作流程。