Validity and reliability of the Modified Sphygmomanometer Test with fixed stabilization for clinical measurement of muscle strength.

作者信息

Silva Brenno B C, Venturato Ana C T, Aguiar Larissa T, Filho Luiz F R M, Faria Christina D C M, Polese Janaine C

机构信息

NeuroEixo, Faculdade Ciências Médicas de Minas Gerais, Belo Horizonte, Minas Gerais, Brazil.

NeuroGroup, Universidade Federal de Minas Gerais, Belo Horizonte, Brazil.

出版信息

J Bodyw Mov Ther. 2019 Oct;23(4):844-849. doi: 10.1016/j.jbmt.2019.05.008. Epub 2019 May 14.

Abstract

OBJECTIVE

To investigate the concurrent validity of the Modified Sphygmomanometer Test (MST) with fixed stabilization, compared to the portable dynamometer, and to verify its test-retest and inter-raterreliability.

METHODS

Methodological study. The muscle strength of the following groups was measured: flexors and extensors of the wrist, of the hip, and of the knee and plantar flexors. The Tycos® aneroid sphygmomanometer and the Microfet 2® dynamometer were used. Stabilization was performed using the Mullingan® belt. Descriptive statistics were performed for characterization of the sample. To determine the validity of the MST with fixed stabilization, comparing it with the portable dynamometer, we investigated the correlation between the measurements obtained with the two instruments using the Pearson correlation coefficient. Intraclass Correlation Coefficient (ICC) was used to investigate inter-examiner and test-retest reliability (α = 0.05).

RESULTS

59 individuals were included (1176 evaluations). A statistically significant correlation of moderate to high magnitude (0.58 ≤ r ≤ 0.81) was observed for concurrent validity of the MST with fixed stabilization versus the portable dynamometer for all muscle groups. Regarding inter-rater reliability, it was observed statistically significant ICC considered excellent to good (0.72 ≤ r ≤ 0.94) for all muscle groups. Regarding test-retest reliability, first evaluator presented good to excellent ICC (0.64 ≤ r ≤ 0.94), while second evaluator presented good to excellent ICC (0.74 ≤ r ≤ 0.96) for all muscle groups.

CONCLUSION

The MST with fixed stabilization is valid and reliable for clinical measurement of muscle strength and can overcome previous limitations reported in the literature.

摘要

目的

研究固定稳定状态下改良血压计测试(MST)与便携式测力计相比的同时效度,并验证其重测信度和评分者间信度。

方法

方法学研究。测量以下几组肌肉的力量:腕部、髋部、膝部的屈肌和伸肌以及足底屈肌。使用泰科斯®无液血压计和Microfet 2®测力计。使用穆林根®束带进行固定。进行描述性统计以描述样本特征。为了确定固定稳定状态下MST的效度,将其与便携式测力计进行比较,我们使用皮尔逊相关系数研究了两种仪器测量结果之间的相关性。组内相关系数(ICC)用于研究评分者间和重测信度(α = 0.05)。

结果

纳入59名个体(1176次评估)。对于所有肌肉组,固定稳定状态下的MST与便携式测力计的同时效度观察到中度至高度的统计学显著相关性(0.58≤r≤0.81)。关于评分者间信度,所有肌肉组的ICC在统计学上被认为是优秀至良好(0.72≤r≤0.94)。关于重测信度,所有肌肉组的第一位评估者的ICC为良好至优秀(0.64≤r≤0.94),而第二位评估者的ICC为良好至优秀(0.74≤r≤0.96)。

结论

固定稳定状态下的MST在临床测量肌肉力量方面是有效且可靠的,并且可以克服文献中报道的先前局限性。