使用前后位 X 线片测量髋臼前倾角的准确性。

Accuracy of planar anteversion measurements using anteroposterior radiographs.

机构信息

Department of Orthopaedic Surgery, Chosun University Hospital, 365 Pilmundae-ro, Dong-gu, Gwangju, 61453, Republic of Korea.

出版信息

BMC Musculoskelet Disord. 2019 Dec 5;20(1):586. doi: 10.1186/s12891-019-2979-0.

Abstract

BACKGROUND

Several methods using simple anteroposterior (AP) radiographs have been suggested for the measurement of anteversion of the cup component after total hip arthroplasty. Herein, we compared six widely used anteversion measurement methods using two different types of AP radiograph, the conventional pelvis AP and hip-centered AP radiographs, to identify the measurement method and the type of radiograph that would provide the highest accuracy and reliability.

METHODS

We developed two custom-made bi-planar anteversion measurement models for the validation test. The models were designed for pelvis AP and hip-centered AP radiographs, respectively. The radiographs were acquired using the inclination angles of both models, changing from 10° to 70° at 10° increments. For each inclination angle, anteversion was changed from 0° to 30° at 5° increments. The measurements were obtained independently by two orthopedic surgeons blinded from each other's measurements, using the methods of 1) Pradhan et al., 2) Lewinnek et al., 3) Widmer et al., 4) Liaw et al., 5) Hassan et al., and 6) Ackland et al. The measurements were repeated after 2 months. The accuracy, compared with that of the reference angle, and intra-observer and inter-observer reliabilities of each method were calculated.

RESULTS

The highest accuracy was found when the method of Liaw et al. was used with hip-centered AP radiographs, which showed a difference of 1.37° ± 1.73 from the reference angle. Moreover, regardless of the type of radiograph, the methods by Pradhan et al., Lewinnek et al., and Liaw et al. showed excellent correlations with the reference anteversion. However, substantial differences were found when the methods by Widmer et al., Hassan et al., and Ackland et al. were used, regardless of the type of radiograph used. When anteversion was measured in an inclination between 30° and 50°, the method of Pradhan et al., when used with pelvis AP radiographs, showed the highest accuracy (1.23° ± 0.92°). We also found no significant difference in anteversions between the measurements made on pelvic and hip-centered AP radiographs. Both interobserver and intraobserver reliabilities were high for all the measurements tested.

CONCLUSIONS

The methods by Pradhan et al., Liaw et al., and Lewinnek et al. may provide relatively accurate anteversion measurements with high reliability, regardless of the type of radiograph.

摘要

背景

已有多种使用简单前后位(AP)射线照相的方法被提出,用于测量全髋关节置换术后杯组件的前倾角。在此,我们比较了使用两种不同类型的 AP 射线照相(常规骨盆 AP 和髋关节中心 AP 射线照相)的六种广泛使用的前倾角测量方法,以确定哪种测量方法和射线照相类型提供最高的准确性和可靠性。

方法

我们为验证试验开发了两个定制的双平面前倾角测量模型。这些模型分别设计用于骨盆 AP 和髋关节中心 AP 射线照相。使用这两个模型的倾斜角度获取射线照相,以 10°的增量从 10°变化到 70°。对于每个倾斜角度,前倾角以 5°的增量从 0°变化到 30°。两名骨科医生独立进行测量,彼此之间不了解对方的测量结果,使用以下方法进行测量:1)Pradhan 等人的方法,2)Lewinnek 等人的方法,3)Widmer 等人的方法,4)Liaw 等人的方法,5)Hassan 等人的方法,和 6)Ackland 等人的方法。两个月后重复测量。计算每种方法的准确性(与参考角度相比)以及观察者内和观察者间的可靠性。

结果

当使用髋关节中心 AP 射线照相和 Liaw 等人的方法时,准确性最高,与参考角度的差异为 1.37°±1.73°。此外,无论射线照相类型如何,Pradhan 等人、Lewinnek 等人和 Liaw 等人的方法均与参考前倾角高度相关。然而,当使用 Widmer 等人、Hassan 等人和 Ackland 等人的方法时,无论使用哪种射线照相类型,都会发现明显的差异。当倾斜角度在 30°至 50°之间时,当使用骨盆 AP 射线照相时,Pradhan 等人的方法显示出最高的准确性(1.23°±0.92°)。我们还发现,在骨盆和髋关节中心 AP 射线照相上测量的前倾角之间没有显著差异。所有测试的测量值的观察者内和观察者间可靠性均较高。

结论

无论射线照相类型如何,Pradhan 等人、Liaw 等人和 Lewinnek 等人的方法都可能提供相对准确且可靠性高的前倾角测量结果。