We evaluated the relationship and the difference between measurements of version angles (VAs) of the acetabular components in total hip arthroplasty taken using 2 different methods. One VA was measured on an anteroposterior radiograph of the hip joint (VAP) and the other on a cross-table lateral radiograph (VCL) in 97 hips after surgery (clinical data) and 6 sawbone pelvic models (model data). There was a positive correlation between VAP and VCL for both data. Mean and standard deviation of the differences (VCL - VAP) between the 2 measurements were 5 degrees +/- 4.2 degrees in clinical data and -0.01 degrees +/- 0.32 degrees in model data. These differences on clinical data should be taken into consideration when comparing VAs in the literature using different measuring methods.