We describe the measurement of bound water T ( ) of cortical bone in vitro and in vivo with a 3D adiabatic inversion recovery ultrashort echo time (IR-UTE) Cones sequence using a clinical 3T scanner.
A series IR-UTE data from 6 repetition times (TRs) with 5 inversion times (TIs) at each TR were acquired from 12 human tibial bone specimens, and data from 4 TRs with 5 TIs at each TR were acquired from the tibial midshafts of 8 healthy volunteers. The pore water nulling point was calculated from exponential fitting of the inversion recovery curve at each TR. Bone specimens and volunteers were then scanned again with the calculated nulling point at each TR. was derived through exponential fitting of data from IR-UTE images acquired at different TRs using the calculated pore water nulling point for each TR.
In vitro pore water nulling TIs were 141.3 ± 11.6, 123.4 ± 8.9, 101.3 ± 6.2, 88.9 ± 5.3, 74.8 ± 4.2, and 59.2 ± 3.9 ms for the 6 TRs of 500, 400, 300, 250, 200, and 150 ms, respectively. In vivo pore water nulling TIs were 132.8 ± 12.8, 110.3 ± 10.0, 80.0 ± 7.2, and 63.9 ± 5.4 ms for the 4 TRs of 400, 300, 200, and 150 ms, respectively. Excellent exponential fitting was achieved for IR-UTE imaging of bound water with pore water nulled at each TR. The mean was 106.9 ± 6.3 ms in vitro and 112.3 ± 16.4 ms in vivo.
Using the 3D IR-UTE Cones with a variable TR/TI approach, of cortical bone was calculated after complete nulling of pore water signals.
我们描述了使用临床3T扫描仪,通过三维绝热反转恢复超短回波时间(IR-UTE)锥体序列在体外和体内测量皮质骨的结合水T( )。
从12个人类胫骨标本中获取了一系列IR-UTE数据,每个重复时间(TR)有6个重复次数,每个TR有5个反转时间(TI);从8名健康志愿者的胫骨中段获取了每个TR有4个重复次数、每个TR有5个TI的数据。通过对每个TR处反转恢复曲线的指数拟合计算孔隙水零点。然后,使用每个TR处计算出的零点对骨标本和志愿者再次进行扫描。通过使用每个TR计算出的孔隙水零点,对在不同TR处采集的IR-UTE图像数据进行指数拟合得出 。
对于500、400、300、250、200和150 ms的6个TR,体外孔隙水零点TI分别为141.3±11.6、123.4±8.9、101.3±6.2、88.9±5.3、74.8±4.2和59.2±3.9 ms。对于400、300、200和150 ms的4个TR,体内孔隙水零点TI分别为132.8±12.8、110.3±10.0、80.0±7.2和63.9±5.4 ms。对于每个TR处孔隙水被归零的结合水的IR-UTE成像,实现了出色的指数拟合。体外平均 为106.9±6.3 ms,体内平均 为112.3±16.4 ms。
使用具有可变TR/TI方法的三维IR-UTE锥体序列,在孔隙水信号完全归零后计算皮质骨的 。