Carpenter D R, Willeke K
Department of Environmental Health, University of Cincinnati, OH 45267-0056.
Am Ind Hyg Assoc J. 1988 Oct;49(10):492-6. doi: 10.1080/15298668891380114.
A noninvasive, fast, inexpensive new fit testing method has been invented which relates the slope of the pressure decay inside a respirator during breath-holding to the fit of the respirator on the wearer's face. The dynamic pressure test has been compared with the conventional aerosol test at different leakage levels. The results of this comparison show that the sensitivity of the dynamic pressure test is similar to that of the aerosol test. The pressure test, however, is independent of leak site and probe location and can be performed on respirators before and after their use.