Harada Yuko, Harada Keiichi, Chin Paul
Internal Medicine, Harada Naika Clinic, Kawasaki, JPN.
Cureus. 2019 Dec 1;11(12):e6268. doi: 10.7759/cureus.6268.
Self-monitoring blood glucose (SMBG) devices have been widely used in medical practice for decades. However, there are certain gaps between SMBG device readings and venous blood glucose levels. Here, 3,532 measurement data points were recorded over 25 years to compare SMBG device readings and venous blood glucose levels. The mean absolute difference (MAD) or the GAP was 10.9 mg/dL, and the mean absolute relative difference (MARD) was 8.3%. The absolute relative difference (ARD) (%) and absolute difference (AD or GAP) (mg/dL) coefficients of variation (CV) (%) of 100% indicate high variance between the capillary BG and venous true BG data. There was a slight skewing of MAD and MARD to the lower body mass index (BMI) side, the higher age side, and the female gender side. There were 41 data points that showed unacceptable gaps of over 50 mg/dL. Such large differences may cause incorrect medications or treatments. All healthcare providers should be aware of the gap between SMBG device readings and venous blood glucose levels.
自我血糖监测(SMBG)设备在医学实践中已广泛使用数十年。然而,SMBG设备读数与静脉血糖水平之间存在一定差距。在此,25年间记录了3532个测量数据点,以比较SMBG设备读数与静脉血糖水平。平均绝对差(MAD)或差距为10.9mg/dL,平均绝对相对差(MARD)为8.3%。100%的绝对相对差(ARD)(%)和绝对差(AD或差距)(mg/dL)变异系数(CV)(%)表明毛细血管血糖与静脉真实血糖数据之间存在高变异性。MAD和MARD在较低体重指数(BMI)一侧、较高年龄一侧和女性一侧存在轻微偏斜。有41个数据点显示出超过50mg/dL的不可接受差距。如此大的差异可能导致用药或治疗错误。所有医疗保健提供者都应意识到SMBG设备读数与静脉血糖水平之间的差距。