Department of Chemistry, University of Waterloo, Waterloo, ON, N2L 3G1, Canada.
Restek Corporation, Bellefonte, PA, 16823, USA.
Anal Bioanal Chem. 2020 Aug;412(21):5067-5076. doi: 10.1007/s00216-019-02367-z. Epub 2020 Jan 6.
Immunosuppressive drugs (ISDs) are primarily administered following solid organ transplant or for treatment of a variety of autoimmune conditions. Their principal function is to suppress the activity of the immune system; however, the levels must be carefully monitored due to adverse effects of over- or underadministration. A technology for rapid quantitative screening, named coated blade spray (CBS), was directly coupled to a triple quadrupole mass spectrometer (MS/MS) to measure the concentration of ISDs (i.e., cyclosporine A, tacrolimus, everolimus, sirolimus) in whole blood samples. We evaluated the stability of replicate measurements over a 10-day period (precision), assessed linearity and limit of quantification, and performed a method comparison against a validated clinical immunoassay (Abbott ARCHITECT). Total interday variation of less than 5% for all target compounds at three different concentrations was achieved. The sensitivity of the method was determined as 0.25, 1, 1, and 2.5 ng/mL for everolimus, sirolimus, tacrolimus, and cyclosporine A, respectively. The concentrations of three immunosuppressive drugs in 284 patient samples (i.e., ~ 95 samples of cyclosporine A, tacrolimus, or sirolimus) obtained using the CBS-MS/MS methodology were compared with concentrations previously quantified on an Abbott ARCHITECT immunoassay system. Our analysis demonstrated significant statistical similarities between both methods. The results demonstrate that CBS-MS/MS is a suitable alternative to conventional methodologies for monitoring of ISDs from whole blood in a clinical setting. Graphical abstract.
免疫抑制剂(ISD)主要在实体器官移植后或用于治疗各种自身免疫性疾病时使用。它们的主要功能是抑制免疫系统的活性;然而,由于给药过多或过少的不良反应,必须仔细监测其水平。一种名为涂层刀片喷雾(CBS)的快速定量筛选技术被直接与三重四极杆质谱仪(MS/MS)耦合,以测量全血样本中 ISD(即环孢素 A、他克莫司、依维莫司、西罗莫司)的浓度。我们评估了在 10 天内重复测量的稳定性(精密度),评估了线性和定量下限,并与经过验证的临床免疫测定法(雅培 ARCHITECT)进行了方法比较。在三个不同浓度下,所有目标化合物的总日内变异小于 5%。该方法的灵敏度分别为 0.25、1、1 和 2.5 ng/mL,用于依维莫司、西罗莫司、他克莫司和环孢素 A。使用 CBS-MS/MS 方法获得的 284 个患者样本(即约 95 个环孢素 A、他克莫司或西罗莫司样本)中三种免疫抑制剂的浓度与以前在雅培 ARCHITECT 免疫测定系统中定量的浓度进行了比较。我们的分析表明两种方法之间存在显著的统计学相似性。结果表明,CBS-MS/MS 是一种替代传统方法的合适方法,可用于临床环境中监测全血中的 ISD。