Suppr超能文献

通过手持测力计测量的髋部力量的测量属性:整个运动范围内的可靠性和有效性。

Measurement properties of hip strength measured by handheld dynamometry: Reliability and validity across the range of motion.

作者信息

Bazett-Jones David M, Squier Kaitlin

机构信息

Department of Athletic Training, University of Toledo, Health & Human Services, Mail Stop 119, 2801 W. Bancroft St, Toledo, OH, 43606-3390, USA.

Department of Physical Therapy, Carroll University, Waukesha, WI, USA.

出版信息

Phys Ther Sport. 2020 Mar;42:100-106. doi: 10.1016/j.ptsp.2020.01.005. Epub 2020 Jan 10.

Abstract

OBJECTIVES

Different angles and positions may influence the reliability and validity of hip strength dynamometry. This study examined the reliability and validity of hip strength measured with belt-stabilized handheld dynamometry (HHD) compared to isokinetic dynamometry (IKD) across multiple angles.

DESIGN

Repeated measures.

SETTING

Biomechanics laboratory.

PARTICIPANTS

Thirty healthy participants completed 3 sessions.

MAIN OUTCOME MEASURES

Hip torque was measured for hip abduction (ABD; -10°, 0°, 20°, and 40°), internal rotation (IR; -30°, -15°, 0°, 15°, and 30°), external rotation (ER; -30°, -15°, 0°, 15°, and 30°), and extension (EXT; -10°, 0°, 30°, 60°, 90°, and 120°).

RESULTS

ABD was reliable (no intratester reliability differences p = 0.409-0.997; ICC = 0.624-0.900) and valid (no HHD-IKD difference p = 0.270-0.946; r = 0.556-0.624) measure. IR had questionable reliability (intratester reliability differences p = 0.001-0.043; ICC = 0.682-0.879) and validity (HHD-IKD differences p < 0.001; r = 0.393-0.771). ER was reliable (no intratester reliability differences p = 0.226-0.937; ICC = 0.731-0.876) but not valid (HHD-IKD differences p < 0.001-0.003; r = 0.194-0.651). EXT showed varying levels of reliability and validity; 30° was both reliable and valid. Peak torque occurred in a lengthened position for ABD, ER, and IR, and at 90° of hip flexion for EXT.

CONCLUSIONS

HHD can be an effective tool for measuring strength under certain circumstances; however, IR/ER testing should be used with caution.

摘要

目的

不同的角度和体位可能会影响髋部力量测力法的可靠性和有效性。本研究比较了在多个角度下,使用带稳定手持式测力计(HHD)与等速测力计(IKD)测量髋部力量的可靠性和有效性。

设计

重复测量。

设置

生物力学实验室。

参与者

30名健康参与者完成了3次测试。

主要观察指标

测量髋关节外展(ABD;-10°、0°、20°和40°)、内旋(IR;-30°、-15°、0°、15°和30°)、外旋(ER;-30°、-15°、0°、15°和30°)和伸展(EXT;-10°、0°、30°、60°、90°和120°)时的髋部扭矩。

结果

ABD测量结果可靠(测试者内可靠性差异p = 0.409 - 0.997;组内相关系数ICC = 0.624 - 0.900)且有效(HHD与IKD差异p = 0.270 - 0.946;r = 0.556 - 0.624)。IR的可靠性存疑(测试者内可靠性差异p = 0.001 - 0.043;ICC = 0.682 - 0.879),有效性也存疑(HHD与IKD差异p < 0.001;r = 0.393 - 0.771)。ER测量结果可靠(测试者内可靠性差异p = 0.226 - 0.937;ICC = 0.731 - 0.876),但无效(HHD与IKD差异p < 0.001 - 0.003;r = 0.194 - 0.651)。EXT的可靠性和有效性水平各不相同;30°时既可靠又有效。ABD、ER和IR的峰值扭矩出现在拉长位置,EXT的峰值扭矩出现在髋关节屈曲90°时。

结论

在某些情况下,HHD可以是测量力量的有效工具;然而,IR/ER测试应谨慎使用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验