Medicina (Kaunas). 2020 Jan 20;56(1):44. doi: 10.3390/medicina56010044.
: the aim of the study was to evaluate the repeatability and reproducibility of optical microangiography (OMAG)-based optical coherence tomography angiography (OCTA) in the optic nerve head (ONH) and radial peripapillary capillary (RPC) perfusion assessment of healthy eyes. : in this observational study, a total of 40 healthy subjects underwent ONH evaluation, using an OMAG-based OCTA system at baseline (T), after 30 min (T), and after 7 days (T). The main outcome measures were the vessel density (VD) and flux index (FI) of the RPCs, as well as peri-papillary retinal nerve fibre layer (pRNFL) thickness. The analysis was performed by two observers independently. The coefficient of repeatability (CR), within the subject coefficient of variation (CVw) and intrasession correlation coefficient (ICC), to evaluate intrasession repeatability of measurements was calculated for each observer. : the high intrasession and intersession repeatability and reproducibility were assessed in the two observers for all three outcome measures. Of note, the CRs for the first and the second observer were 0.011 (95% confidence interval (CI) 0.009-0.014) and 0.016 (95% CI 0.013-0.020) for FI, 0.016 (95% CI 0.013-0.021) and 0.017 (95% CI 0.014-0.021) for VD, and 2.400 (95% CI 1.948-3.092) and 3.732 (95% CI 3.064-4.775) for pRNFL thickness, respectively. The agreement between them was excellent for pRNFL assessment and very good for FI and VD. : OCTA has a great potential in the accurate assessment of ONH and peri-papillary microcirculation. It allows for repeated and reproducible measurements without multiple scans-related bias, thus guaranteeing an independent operator analysis with good reproducibility and repeatability.
本研究旨在评估基于光学微血管造影(OMAG)的光学相干断层扫描血管造影(OCTA)在健康人视神经头(ONH)和视盘周围毛细血管(RPC)灌注评估中的重复性和可再现性。
在这项观察性研究中,共有 40 名健康受试者接受了基于 OMAG 的 OCTA 系统的 ONH 评估,基线(T)时、30 分钟(T)时和 7 天后(T)时。主要观察指标为 RPC 的血管密度(VD)和血流指数(FI)以及视盘周围视网膜神经纤维层(pRNFL)厚度。分析由两位观察者独立进行。为了评估每位观察者测量的日内重复性,计算了受试者内变异系数(CVw)内的重复性系数(CR)和组内相关系数(ICC)。
两位观察者对所有三种测量结果均评估了高日内和日间的重复性和可再现性。值得注意的是,对于 FI,第一和第二位观察者的 CR 分别为 0.011(95%置信区间[CI]0.009-0.014)和 0.016(95% CI 0.013-0.020),对于 VD,分别为 0.016(95% CI 0.013-0.021)和 0.017(95% CI 0.014-0.021),对于 pRNFL 厚度,分别为 2.400(95% CI 1.948-3.092)和 3.732(95% CI 3.064-4.775)。对于 pRNFL 评估,他们之间的一致性非常好,而对于 FI 和 VD,则非常好。
OCTA 在准确评估 ONH 和视盘周围微循环方面具有巨大潜力。它允许进行重复和可再现的测量,而不会产生与多次扫描相关的偏差,从而保证了具有良好可重复性和重复性的独立操作者分析。