This is the first study to validate the repeatability of objective measurements of amplitude of accommodation (AA) using the TONOREF III (NIDEK Co., Ltd., Japan), which can measure the AA in 30 seconds.
To evaluate the repeatability of objective measurements of AA using the TONOREF III and explored the association between objectively measured AA and factors, including age, sex, spherical equivalent, baseline pupil size and pupil size change during accommodation.
This cross-sectional study recruited 35 healthy subjects aged 26 to 52 years. The Bland-Altman method and intraclass correlation coefficients (ICCs) were used to assess the repeatability of TONOREF III measurements. The relationships between AA and age, sex, baseline pupil size, changes in pupil size (ΔPS) during accommodation were analyzed using linear regression for univariate and multivariate analysis.
The mean difference in AA (ΔAA) between two sessions of TONOREF III measurements was 0.23 D (95% CI: -1.07 to +1.53 D), while no significant correlation was found between the mean and ΔAA (p = .14). The ICCs of the TONOREF III was 0.96. Age, sex, and ΔPS during accommodation were significant factors affecting the AA in multivariate analysis.
The repeatability of objective AA measurements using the TONOREF III was good. Measuring AA using the TONOREFTM III in clinical practice is feasible.
这是第一项使用 TONOREF III(日本尼德克株式会社)验证调节幅度(AA)客观测量重复性的研究,该仪器可在 30 秒内测量 AA。
评估使用 TONOREF III 进行 AA 客观测量的重复性,并探讨客观测量的 AA 与包括年龄、性别、等效球镜、基线瞳孔大小和调节时瞳孔大小变化等因素之间的关系。
这项横断面研究招募了 35 名年龄在 26 至 52 岁的健康受试者。采用 Bland-Altman 方法和组内相关系数(ICCs)评估 TONOREF III 测量的重复性。使用单变量和多变量分析的线性回归分析 AA 与年龄、性别、基线瞳孔大小、调节时瞳孔大小变化(ΔPS)之间的关系。
两次 TONOREF III 测量的 AA(ΔAA)平均值差异为 0.23 D(95%CI:-1.07 至 +1.53 D),但平均值与 ΔAA 之间无显著相关性(p =.14)。TONOREF III 的 ICC 为 0.96。年龄、性别和调节时的 ΔPS 是多变量分析中影响 AA 的显著因素。
使用 TONOREF III 进行客观 AA 测量的重复性良好。在临床实践中使用 TONOREFTM III 测量 AA 是可行的。