Camardella Leonardo T, Ongkosuwito Edwin M, Penning E Willemijn, Kuijpers-Jagtman Anne Marie, Vilella Oswaldo V, Breuning K Hero
Department of Orthodontics, School of Dentistry, Universidade Federal Fluminense, Niterói, Brazil.
Department of Orthodontics and Craniofacial Biology, Radboud University Medical Center, Nijmegen, The Netherlands.
Korean J Orthod. 2020 Jan;50(1):13-25. doi: 10.4041/kjod.2020.50.1.13. Epub 2020 Jan 22.
The aim of this study was to compare the accuracy and reliability of measurements performed using two different software programs on digital models generated using two types of plaster model scanners (a laser scanner and a computed tomography [CT] scanner).
Thirty plaster models were scanned with a 3Shape laser scanner and with a Flash CT scanner. Two examiners performed measurements on plaster models by using digital calipers and on digital models by using Ortho Analyzer (3Shape) and Digimodel® (OrthoProof) software programs. Forty-two measurements, including tooth diameter, crown height, overjet, overbite, intercanine and intermolar distances, and sagittal relationship, were obtained.
Statistically significant differences were not found between the plaster and digital model measurements (ANOVA); however, some discrepancies were clinically relevant. Plaster and digital model measurements made using the two scanning methods showed high intraclass coefficient correlation values and acceptable 95% limits of agreement in the Bland-Altman analysis. The software used did not influence the accuracy of measurements.
Digital models generated from plaster casts by using laser and CT scanning and measured using two different software programs are accurate, and the measurements are reliable. Therefore, both fabrication methods and software could be used interchangeably.
本研究旨在比较使用两种不同软件程序对两种类型的石膏模型扫描仪(激光扫描仪和计算机断层扫描[CT]扫描仪)生成的数字模型进行测量的准确性和可靠性。
用3Shape激光扫描仪和Flash CT扫描仪对30个石膏模型进行扫描。两名检查者分别使用数字卡尺在石膏模型上以及使用Ortho Analyzer(3Shape)和Digimodel®(OrthoProof)软件程序在数字模型上进行测量。共获得42项测量数据,包括牙齿直径、冠高、覆盖、覆合、尖牙间和磨牙间距离以及矢状关系。
石膏模型和数字模型测量结果之间未发现统计学上的显著差异(方差分析);然而,一些差异具有临床相关性。使用两种扫描方法进行的石膏模型和数字模型测量在组内相关系数分析中显示出较高的组内相关系数值和可接受的95%一致性界限。所使用的软件并未影响测量的准确性。
通过激光和CT扫描从石膏模型生成并使用两种不同软件程序测量的数字模型是准确的,测量结果是可靠的。因此,两种制作方法和软件均可互换使用。