在超重或肥胖患者中使用150 kVp的光谱成形进行胸部CT筛查:与120 kVp方案和100 kVp方案的光谱成形相比。

Chest CT screening in patients with overweight or obesity using spectral shaping at 150 kVp: compared with 120 kVp protocol and spectral shaping at 100 kVp protocol.

作者信息

Li Jianwen, Mai Zhifeng, Zhang Zhihong, Cui Jiamang, Yang Mingjie, Ma Xia, Wang Yan

机构信息

Department of Radiology, The Second People's Hospital of Shizuishan, NO.246 West Youyi Street, Shizuishan, 753000, Ningxia, China.

Department of Pharmacy, The First People's Hospital of Shizuishan, NO.1 Kangle Road, Shizuishan, 753000, Ningxia, China.

出版信息

Jpn J Radiol. 2020 May;38(5):451-457. doi: 10.1007/s11604-020-00925-5. Epub 2020 Feb 11.

Abstract

PURPOSE

To evaluate the image quality (IQ) and the figure of merit (FoM) of chest CT screening in patients with overweight or obesity using a tin filter for spectral shaping at 150 kVp.

MATERIALS AND METHODS

Patients with overweight or obesity (N = 150, body mass index ≥ 26 kg/m) with indications for chest CT screening were prospectively enrolled and randomly divided into three groups: 120 kVp group (standard radiation dose/tube voltage, 120 kVp/CT volume does index, 4.68 mGy); Sn100 kVp group (1/10th radiation dose level/100 kVp with a tin filter/0.47 mGy); Sn150 kVp group (1/2th radiation dose level/150 kVp with a tin filter/2.34 mGy). IQ and FoMs were evaluated and compared among the three groups.

RESULTS

Image noise, signal-to-noise ratios and subjective IQ scores were significantly higher in the Sn150 kVp group than those in the Sn100 kVp group (all p < 0.05), but were not significantly different with those in the 120 kVp group. FoMs in the Sn150 kVp group were significantly higher than those in the 120 kVp group (all p < 0.05), but showed no statistical difference with those in the Sn100 kVp group.

CONCLUSIONS

Compared with scanning at 120 kVp, chest CT screening performed at 150 kVp with spectral shaping substantially reduces the radiation dose in overweight and obese patients while maintaining IQ.

摘要

目的

使用锡滤过器在150 kVp下进行能谱塑形,评估超重或肥胖患者胸部CT筛查的图像质量(IQ)和品质因数(FoM)。

材料与方法

前瞻性纳入有胸部CT筛查指征的超重或肥胖患者(N = 150,体重指数≥26 kg/m²),并随机分为三组:120 kVp组(标准辐射剂量/管电压,120 kVp/CT容积剂量指数,4.68 mGy);Sn100 kVp组(1/10辐射剂量水平/100 kVp并使用锡滤过器/0.47 mGy);Sn150 kVp组(1/2辐射剂量水平/150 kVp并使用锡滤过器/2.34 mGy)。对三组的IQ和FoM进行评估并比较。

结果

Sn150 kVp组的图像噪声、信噪比和主观IQ评分显著高于Sn100 kVp组(均p < 0.05),但与120 kVp组无显著差异。Sn150 kVp组的FoM显著高于120 kVp组(均p < 0.05),但与Sn100 kVp组无统计学差异。

结论

与120 kVp扫描相比,在150 kVp下进行能谱塑形的胸部CT筛查在超重和肥胖患者中可大幅降低辐射剂量,同时保持图像质量。