100kVp 锡过滤下的儿科胸部 CT:在可比辐射剂量下与 70kVp 成像的图像质量比较。

Pediatric chest computed tomography at 100 kVp with tin filtration: comparison of image quality with 70-kVp imaging at comparable radiation dose.

机构信息

Department of Thoracic Imaging, Hospital Calmette, Blvd Jules Leclercq, 59037, Lille cedex, France.

Santé Publique: épidémiologie et qualité des soins, University of Lille, CHU Lille, Lille, France.

出版信息

Pediatr Radiol. 2020 Feb;50(2):188-198. doi: 10.1007/s00247-019-04543-w. Epub 2019 Dec 18.

Abstract

BACKGROUND

Radiation dose reduction is a primary objective in pediatric populations owing to the well-known risks of radiation-induced cancers. Low-energy photons participate in the radiation dose without significantly contributing to image formation. Their suppression by means of tin filtration should decrease the image noise, anticipating a subsequent application to dose saving.

OBJECTIVE

To investigate the level of noise reduction achievable with tin (Sn) filtration at 100 kVp for chest computed tomography (CT) in comparison with a standard scanning mode at 70 kVp with comparable radiation dose.

MATERIALS AND METHODS

Fifty consecutive children (Group 1) underwent non-contrast chest CT examinations on a third-generation dual-source CT system at tin-filtered 100 kVp and pitch 2. The tube-current time product (mAs) was adjusted to maintain the predicted dose length product (DLP) value at 70 kVp for the respective patient. Each child was then paired by weight and age to a child scanned at 70 kVp on the same CT unit (Group 2); Group 2 patients were consecutive patients, retrospectively selected from our database of children prospectively scanned at 70 kVp. Objective and subjective image quality were compared between the two groups of patients to investigate the overall image quality and level of noise reduction that could be subsequently achievable with tin filtration in clinical practice.

RESULTS

The mean image noise was significantly lower in Group 1 compared to Group 2 when measured in the air (P<0.0001) and inside the aorta (P<0.001). The mean noise reduction was 21.6% (standard deviation [SD] 16.1) around the thorax and 12.0% (SD 32.7) inside the thorax. There was no significant difference in subjective image quality of lung and mediastinal images with excellent overall subjective scores in both groups.

CONCLUSION

At comparable radiation dose, the image noise was found to be reduced by 21.6% compared to the 70-kVp protocol, providing basis for dose reduction without altering image quality in further investigations.

摘要

背景

由于众所周知的辐射致癌风险,降低儿科人群的辐射剂量是主要目标。低能光子参与辐射剂量的形成,但对图像形成的贡献不大。通过锡滤过抑制它们应该会降低图像噪声,从而为后续的剂量节省应用提供依据。

目的

在与 70 kVp 标准扫描模式相比辐射剂量相当的情况下,研究 100 kVp 时锡(Sn)滤过对胸部 CT(CT)的噪声降低程度。

材料和方法

50 名连续的儿童(第 1 组)在第三代双源 CT 系统上进行非对比胸部 CT 检查,管电压为 100 kVp 并采用 2 的螺距。管电流时间乘积(mAs)被调整,以维持各自患者 70 kVp 时的预测剂量长度乘积(DLP)值。然后,根据体重和年龄将每个儿童与在同一 CT 机上以 70 kVp 扫描的儿童配对(第 2 组);第 2 组患者是从我们的 70 kVp 前瞻性扫描儿童数据库中回顾性选择的连续患者。比较两组患者的客观和主观图像质量,以研究在临床实践中锡滤过随后可实现的整体图像质量和噪声降低水平。

结果

与第 2 组相比,第 1 组在空气中(P<0.0001)和主动脉内(P<0.001)测量的平均图像噪声明显更低。胸部周围的平均噪声降低为 21.6%(标准差 [SD] 16.1),胸部内为 12.0%(SD 32.7)。两组的肺和纵隔图像的主观图像质量没有显著差异,两组的总体主观评分均为优秀。

结论

在可比辐射剂量下,与 70 kVp 方案相比,图像噪声降低了 21.6%,为进一步研究中不改变图像质量的情况下降低剂量提供了依据。