Merle Geraldine, Comeau-Gauthier Marianne, Tayari Vahid, Kezzo Mohamad Nizar, Kasem Chrouk, Al-Kabraiti Faisal, Laverdiere Carl, Xereas George, Harvey Edward J
McGill University Health Center Research Institute, Montreal General Hospital, 1650 Cedar Ave, Montreal, QC H3G1A4.
Department of Surgery, McGill University, Montreal General Hospital, 1650 Cedar Ave, Montreal, QC H3G1A4.
Mil Med. 2020 Jan 7;185(Suppl 1):77-81. doi: 10.1093/milmed/usz305.
Acute compartment syndrome (ACS) is a well-recognized and common emergency. Undiagnosed ACS leads to muscle necrosis, limb contracture, intractable pain, and may even result in amputation.
Three devices (Synthes, Stryker, and MY01) were compared in a pre-clinical rat abdominal compartment syndrome simulation. Simultaneous measurements of intracompartmental pressures allowed concurrent comparison among all devices.
Large variations from the reference values are seen with the Synthes and Stryker devices. Variances are large in these two devices even under ideal conditions. The MY01 device was the truest indicator of reference pressure in this ACS model (over 600% more accurate).
The MY01 device was the most accurate device in tracking pressure changes in this rat model of abdominal compartment syndrome.