Ophthalmology Department, Kayseri City Education and Research Hospital, Building of Pschiatry 1st Floor Ophthalmology Service, Kocasinan, Kayseri, Turkey.
Int Ophthalmol. 2020 Jun;40(6):1481-1485. doi: 10.1007/s10792-020-01315-6. Epub 2020 Feb 19.
To define diurnal changes in anterior segment parameters of keratoconus patients by using Scheimpflug imaging.
All keratoconus patients had corneal topography measurements 3 times a day (around 09:00 AM, 13:00 PM and 17:00 PM) by the same experienced operator. Three consecutive scans in each measurement session were obtained. The average of three measurements was used for analysis. A repeated measures analysis of variance with one within-subject factor (time of day) was carried out to assess diurnal variation.
This study included 26 eyes of 26 patients with keratoconus. Significant diurnal variation was found in CCT and TCT measurements (P < 0.01, and P < 0.01, respectively). The mean amplitudes of change in CCT and TCT were 4.2 ± 1 µm (95% CI: 1.7-6.8 µm) and 4.1 ± 1.1 µm (95% CI: 1.2-8.0 µm), while the mean of CCT and TCT were 462.4 ± 34.5 µm (95% CI: 448.4-476.6 µm) and 452.9 ± 6.6 µm (95% CI: 439.3-466.5 µm), respectively. The maximum value and the lowest value were observed at 09:00 AM and at 17:00 PM, respectively. Other parameters except CCT and TCT did not show diurnal variation.
Statistically significant but clinically insignificant diurnal variation in CCT and TCT measurements of the keratoconus patients were detected over an 8-hour period (09:00 AM-17:00 PM).
通过 Scheimpflug 成像来定义圆锥角膜患者前节参数的昼夜变化。
同一位经验丰富的操作者在一天中的三个时间点(上午 9 点、下午 1 点和下午 5 点)对所有圆锥角膜患者进行 3 次角膜地形图测量。在每次测量中获得 3 个连续的扫描。分析时使用三次测量的平均值。采用单因素重复测量方差分析评估昼夜变化。
这项研究包括 26 例 26 只眼的圆锥角膜患者。CCT 和 TCT 测量值存在显著的昼夜变化(P<0.01 和 P<0.01)。CCT 和 TCT 变化的平均幅度分别为 4.2±1μm(95%CI:1.7-6.8μm)和 4.1±1.1μm(95%CI:1.2-8.0μm),而 CCT 和 TCT 的平均值分别为 462.4±34.5μm(95%CI:448.4-476.6μm)和 452.9±6.6μm(95%CI:439.3-466.5μm)。最大值和最小值分别出现在上午 9 点和下午 5 点。除 CCT 和 TCT 外,其他参数均无昼夜变化。
在 8 小时(上午 9 点至下午 5 点)期间,检测到圆锥角膜患者 CCT 和 TCT 测量值存在统计学上显著但临床上无意义的昼夜变化。