J Cataract Refract Surg. 2011 Oct;37(10):1809-16. doi: 10.1016/j.jcrs.2011.04.033.
To assess the repeatability of anterior segment measurements performed by a Scheimpflug camera combined with Placido corneal topography (Sirius) in unoperated, post-refractive surgery, and keratoconus eyes.
Private clinical ophthalmology practice.
Evaluation of diagnostic test or technology.
Three consecutive scans were acquired for each eye. The following parameters were evaluated: simulated keratometry, posterior corneal power, mean pupil power (ie, corneal power assessed by ray tracing through the anterior and posterior corneal surfaces), corneal asphericity, thinnest and apex corneal thickness, aqueous depth, anterior chamber volume, and corneal spherical aberration. Repeatability was assessed using test-retest variability, the coefficient of variation, and the intraclass correlation coefficient (ICC).
Sixty-four unoperated eyes, 17 eyes that had myopic excimer laser surgery, and 13 eyes with keratoconus were analyzed. High repeatability was achieved for most parameters in the 3 groups, with an ICC higher than 0.99 for all measurements except posterior corneal power and mean pupil power in keratoconus (ICC, 0.868 and 0.976, respectively), anterior and posterior asphericity in normal eyes (ICC, 0.904 and 0.977, respectively), and spherical aberration in normal eyes (ICC, 0.806), post-refractive surgery eyes (ICC, 0.980), and keratoconus eyes (ICC, 0.981).
The anterior segment measurements provided by the new Scheimpflug camera-Placido corneal topography system were highly repeatable and can be relied on in clinical routine and for research purposes.
No author has a financial or proprietary interest in any material or method mentioned.
评估应用 Scheimpflug 相机联合 Placido 角膜地形图(Sirius)对未行手术、行屈光手术后和圆锥角膜患者眼前节进行测量的重复性。
私人临床眼科诊所。
诊断测试或技术评估。
对每只眼连续采集 3 次扫描。评估以下参数:模拟角膜曲率、后表面角膜曲率、平均瞳孔力(即通过前、后角膜表面光线追踪评估的角膜曲率)、角膜非球面性、最薄和顶点角膜厚度、房水深度、前房容积和角膜球差。采用测试-再测试变异性、变异系数和组内相关系数(ICC)评估重复性。
共分析了 64 只未手术眼、17 只行近视准分子激光手术眼和 13 只圆锥角膜眼。在 3 组中,大多数参数均具有高度重复性,除圆锥角膜眼的后表面角膜曲率和平均瞳孔力(ICC 分别为 0.868 和 0.976)、正常眼的前、后表面非球面性(ICC 分别为 0.904 和 0.977)和正常眼的球差(ICC 为 0.806)、行屈光手术后眼(ICC 为 0.980)和圆锥角膜眼(ICC 为 0.981)外,所有测量的 ICC 均高于 0.99。
新型 Scheimpflug 相机-Placido 角膜地形图系统提供的眼前节测量重复性高,可在临床常规和研究中应用。