Departments of Neurosurgery and Clinical and Experimental Medicine, Linköping University, Linköping, Sweden,
Department of Biomedical Engineering, Linköping University, Linköping, Sweden.
Stereotact Funct Neurosurg. 2020;98(1):55-61. doi: 10.1159/000505708. Epub 2020 Feb 20.
Optics can be used for guidance in deep brain stimulation (DBS) surgery. The aim was to use laser Doppler flowmetry (LDF) to investigate the intraoperative optical trajectory along the ventral intermediate nucleus (VIM) and zona incerta (Zi) regions in patients with essential tremor during asleep DBS surgery, and whether the Zi region could be identified.
A forward-looking LDF guide was used for creation of the trajectory for the DBS lead, and the microcirculation and tissue greyness, i.e., total light intensity (TLI) was measured along 13 trajectories. TLI trajectories and the number of high-perfusion spots were investigated at 0.5-mm resolution in the last 25 mm from the targets.
All implantations were done without complications and with significant improvement of tremor (p < 0.01). Out of 798 measurements, 12 tissue spots showed high blood flow. The blood flow was significantly higher in VIM than in Zi (p < 0.001). The normalized mean TLI curve showed a significant (p < 0.001) lower TLI in the VIM region than in the Zi region.
Zi DBS performed asleep appears to be safe and effective. LDF monitoring provides direct in vivomeasurement of the microvascular blood flow in front of the probe, which can help reduce the risk of hemorrhage. LDF can differentiate between the grey substance in the thalamus and the transmission border entering the posterior subthalamic area where the tissue consists of more white matter tracts.
光学可用于深部脑刺激(DBS)手术中的引导。目的是使用激光多普勒流量测量法(LDF)来研究在睡眠 DBS 手术中特发性震颤患者的腹侧中间核(VIM)和未定带(Zi)区域的术中光学轨迹,以及是否可以识别 Zi 区域。
使用前瞻性 LDF 引导器为 DBS 导联创建轨迹,并测量 13 条轨迹上的微循环和组织灰度,即总光强度(TLI)。在距离靶点最后 25mm 处,以 0.5mm 的分辨率研究 TLI 轨迹和高灌注点的数量。
所有植入均无并发症,并使震颤明显改善(p<0.01)。在 798 次测量中,有 12 个组织点显示出高血流。VIM 中的血流明显高于 Zi(p<0.001)。归一化平均 TLI 曲线显示 VIM 区域的 TLI 明显低于 Zi 区域(p<0.001)。
在睡眠中进行 Zi DBS 似乎是安全有效的。LDF 监测提供了探头前方微血管血流的直接活体测量,这有助于降低出血风险。LDF 可以区分丘脑的灰质和进入后苍白球下区的传导边界,后者的组织包含更多的白质束。