Suppr超能文献

使用实时光纤共聚焦显微镜对心脏传导组织区域进行术中定位:首例人体试验。

Intraoperative localization of cardiac conduction tissue regions using real-time fibre-optic confocal microscopy: first in human trial.

作者信息

Kaza Aditya K, Mondal Abhijit, Piekarski Breanna, Sachse Frank B, Hitchcock Robert

机构信息

Department of Cardiac Surgery, Boston Children's Hospital, Harvard Medical School, Boston, MA, USA.

Department of Biomedical Engineering, University of Utah, Salt Lake City, UT, USA.

出版信息

Eur J Cardiothorac Surg. 2020 Aug 1;58(2):261-268. doi: 10.1093/ejcts/ezaa040.

Abstract

OBJECTIVES

The aim of this study was to evaluate the feasibility and safety of fibre-optic confocal microscopy (FCM) using fluorescein sodium dye for the intraoperative location of conduction tissue regions during paediatric heart surgery.

METHODS

The pilot study included 6 patients undergoing elective surgery for the closure of isolated secundum atrial septal defect aged 30 days to 21 years. FCM imaging was integrated within the normal intraoperative protocol for atrial septal defect repair. Fluorescein sodium dye was applied on the arrested heart. FCM images were acquired at the atrioventricular node region, sinus node region and right ventricle (RV). Total imaging time was limited to 3 min. Any adverse events related to the study were recorded and analysed. Subjects received standard postoperative care. Trained reviewers (n = 9) classified, de-identified and randomized FCM images (n = 60) recorded from the patients as presenting striated, reticulated or indistinguishable microstructures. The reliability of reviewer agreement was assessed using Fleiss' kappa.

RESULTS

The FCM imaging instruments were integrated effectively into the cardiac surgery operating room. All adverse events found in the study were deemed expected and not related to FCM imaging. Reticulated myocardial microstructures were found during FCM imaging at atrioventricular node and sinus node regions, while striated microstructures were observed in RV. Reliability of agreement of reviewers classifying the FCM images was high (Fleiss' kappa: 0.822).

CONCLUSIONS

FCM using fluorescein sodium dye was found to be safe for use during paediatric heart surgery. The study demonstrates the potential for FCM to be effective in identifying conduction tissue regions during congenital heart surgery.

CLINICAL TRIAL REGISTRATION NUMBER

NCT03189134.

摘要

目的

本研究旨在评估使用荧光素钠染料的光纤共聚焦显微镜(FCM)在小儿心脏手术中定位传导组织区域的可行性和安全性。

方法

该初步研究纳入了6例年龄在30天至21岁之间接受择期手术治疗孤立性继发孔房间隔缺损封堵术的患者。FCM成像被纳入房间隔缺损修复的正常术中流程。荧光素钠染料应用于停跳的心脏。在房室结区域、窦房结区域和右心室(RV)采集FCM图像。总成像时间限制为3分钟。记录并分析与该研究相关的任何不良事件。受试者接受标准的术后护理。训练有素的评审员(n = 9)将从患者记录的FCM图像(n = 60)进行分类、去识别并随机分组,这些图像呈现为条纹状、网状或无法区分的微观结构。使用Fleiss卡方评估评审员一致性的可靠性。

结果

FCM成像仪器有效地整合到了心脏手术手术室中。该研究中发现的所有不良事件均被认为是预期的,且与FCM成像无关。在FCM成像过程中,在房室结和窦房结区域发现了网状心肌微观结构,而在右心室观察到了条纹状微观结构。评审员对FCM图像分类的一致性可靠性很高(Fleiss卡方:0.822)。

结论

发现使用荧光素钠染料的FCM在小儿心脏手术中使用是安全的。该研究证明了FCM在先天性心脏手术中识别传导组织区域方面有效的潜力。

临床试验注册号

NCT03189134。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/82c5/7594189/67116e0f5756/ezaa040f3.jpg