Department of Radiology, Ohio State University College of Medicine, 395 W 12th Ave, Columbus, OH 43210.
Department of Radiology, Ohio State University College of Medicine, 395 W 12th Ave, Columbus, OH 43210.
Acad Radiol. 2021 Jan;28(1):96-105. doi: 10.1016/j.acra.2019.12.027. Epub 2020 Feb 21.
The purpose of this study was to investigate the potential effects of patient size and radiation dose on the accuracy of iodine quantification using dual-source dual-energy computed tomography (CT).
Three phantoms representing different patient sizes were constructed, containing iodine inserts with concentrations from 0 to 20 mg/ml. Dual-energy CT scans were performed at six dose levels from 2 to 30 mGy. Iodine concentrations were measured using a three-material-decomposition algorithm and their accuracy was assessed.
In a small phantom, iodine quantification was accurate and consistent at all dose levels. In a medium phantom, minor underestimations were observed, and the results were consistent except at low dose. In the large phantom, more significant underestimation of iodine concentration was observed at higher doses (≥15 mGy), which was attributed to the beam-hardening effect. At lower doses, increasing upward bias was observed in the CT number, leading to significant overestimations of both iodine concentration and fat fraction, which was attributed to the photon-starvation effect. The severity of the latter effect was determined by mA instead of mAs, suggesting that the electronic noise, rather than the quantum noise, was responsible for the bias. Using higher kVp for the low-energy tube was found to alleviate these effects.
Reliable iodine quantification can be achieved using dual-source CT, but the result can be affected by patient size and dose rate. In large patients, biases may occur due to the beam-hardening and the photon-starvation effects, in which case higher dose rate and higher kVp are recommended to minimize these effects.
本研究旨在探讨使用双源双能 CT 时患者体型和辐射剂量对碘定量准确性的潜在影响。
构建了三个代表不同体型的体模,其中包含碘浓度从 0 至 20mg/ml 的碘插入物。在 2 至 30mGy 的六个剂量水平上进行双能 CT 扫描。使用三物质分解算法测量碘浓度,并评估其准确性。
在小体模中,在所有剂量水平下碘定量均准确且一致。在中体模中,观察到轻微低估,结果一致,除了在低剂量下。在大体模中,在较高剂量(≥15mGy)下观察到碘浓度的更显著低估,这归因于束硬化效应。在较低剂量下,观察到 CT 值向上的偏移增加,导致碘浓度和脂肪分数的显著高估,这归因于光子饥饿效应。后者的严重程度由 mA 而不是 mAs 决定,表明是电子噪声而不是量子噪声导致了偏差。发现使用高能管可以减轻这些影响。
使用双源 CT 可以实现可靠的碘定量,但结果可能受到患者体型和剂量率的影响。在大患者中,由于束硬化和光子饥饿效应可能会出现偏差,在这种情况下,建议使用更高的剂量率和更高的 kVp 以最小化这些影响。