Nguyen Callistus M, Baydush Alan H, Ververs James D, Isom Scott, Able Charles M, Munley Michael T
Biomedical Engineering, Virginia Tech - Wake Forest School of Biomedical Engineering and Sciences, 575 N. Patterson Ave, Winston-Salem, NC 27101, United States.
Radiation Oncology, Wake Forest School of Medicine, Medical Center Boulevard, Winston-Salem, NC 27157, United States.
Tech Innov Patient Support Radiat Oncol. 2018 May 29;7:6-10. doi: 10.1016/j.tipsro.2018.05.001. eCollection 2018 Sep.
The purpose of this study was to determine if medical linear accelerators (linac) produced by the same manufacturer exhibit operational consistency within their subsystems and components. Two linacs that were commissioned together and installed at the same facility were monitored. Each machine delivered a daily robust quality assurance (QA) irradiation. Linacs and their components operate consistently, but have different operational parameter levels even when produced by the same manufacturer and commissioned in series. These findings have implications on the feasibility of true clinical beam matching.