30MHz 和 75MHz 高频超声与组织病理学检测基底细胞癌浸润深度的对比研究。

Basal cell carcinoma invasion depth determined with 30 and 75 MHz high-frequency ultrasound and histopathology - a comparative study.

机构信息

M.F. Vladimirsky Moscow Regional Scientific Research Clinical Institute.

Academy of Postgraduate Education under the Federal State Budgetary Institution Federal State Budgetary Unit "Federal Scientific and Clinical Center for Specialized Medical Assistance and Medical Technologies of the Federal Medical Biological Agency".

出版信息

Med Ultrason. 2020 Mar 1;22(1):31-36. doi: 10.11152/mu-2233.

Abstract

AIM

To compare the depth spread of basal cell carcinoma (BCC) measured by histological examination and high-frequency ultrasound (HFUS) imaging with 30-MHz and 75-MHz probes.

MATERIALS AND METHODS

HFUS skin imaging was used to examine 27 BCCs. A specialized high-resolution digital ultrasound imaging system DUB (TPM GmbH, Germany) with 75-MHz and 30-MHz probes was used. After HFUS scanning, the BCCs biopsy samples were collected by punch biopsy or surgical excision for the morphological examination. Based on the histomorphology results obtained, the tumors were divided into thin (≤1 mm invasion depth) and thick (>1 mm invasion depth). Each BCC spread depth was measured during the HFUS examination with 75-MHz and 30-MHz ultrasound probes and morphological examination.

RESULTS

Thin BCCs average invasiondepth measured histologically was 0.494±0.212 mm. Its average depth obtained with HFU examination with 75-MHz and 30-MHz probes was 0.591±0.265 and 0.734±0.123 mm, respectively. High, statistically significant correlation betweenthe histological and 75 MHz HFU measurements was obtained (r=0.870). The correlation was weak (r=0.290) when using a 30 MHz transducer. The average thick BCC invasion depth values obtained with the histological examination and 30 MHz HFUS scanning was 1.845±0.718 mm and 1.995±0.699 mm, respectively. High, statistically significant (r=0.951) correlation between the thick BCC spread depth measured with 30 MHz transducer and histomorphological examination was obtained.

CONCLUSIONS

In cases of BCCs with thickness of ≤1 mm, there was a high correlation (r=0.870) of the tumor spread depth between micromorphological measurements and the results obtained using a 75 MHz transducer and in cases of BCCs with thickness of >1 mm, a very high correlation (r=0.951) of the tumor spread depth was observed between histomorphometry and30 MHz transducer measurements.

摘要

目的

比较组织学检查和高频超声(HFUS)成像测量基底细胞癌(BCC)的深度分布,所用探头频率分别为 30MHz 和 75MHz。

材料与方法

应用 HFUS 皮肤成像检查 27 例 BCC。采用专门的高分辨率数字超声成像系统 DUB(德国 TPM 公司),探头频率分别为 75MHz 和 30MHz。HFUS 扫描后,采用环钻或外科切除法采集 BCC 活检样本,进行形态学检查。根据组织形态学结果,将肿瘤分为薄型(浸润深度≤1mm)和厚型(浸润深度>1mm)。在 HFUS 检查过程中,用 75MHz 和 30MHz 超声探头和形态学检查分别测量每个 BCC 的扩散深度。

结果

组织学检查显示,薄型 BCC 的平均浸润深度为 0.494±0.212mm。HFUS 检查中,用 75MHz 和 30MHz 探头分别测量的平均深度为 0.591±0.265mm 和 0.734±0.123mm。两种方法测量结果之间高度相关(r=0.870)。用 30MHz 探头时,相关性较弱(r=0.290)。组织学检查和 30MHz HFUS 扫描分别测得的厚型 BCC 平均浸润深度值为 1.845±0.718mm 和 1.995±0.699mm。两种方法测量厚型 BCC 扩散深度之间高度相关(r=0.951)。

结论

对于厚度≤1mm 的 BCC,肿瘤扩散深度在微形态学测量与 75MHz 探头结果之间具有高度相关性(r=0.870),而对于厚度>1mm 的 BCC,肿瘤扩散深度在组织形态学与 30MHz 探头测量之间具有极高相关性(r=0.951)。