用于青光眼手术的、连接到180度巩膜外植体上的长克鲁平-丹佛阀门植入物。

A long Krupin-Denver valve implant attached to a 180 degrees scleral explant for glaucoma surgery.

作者信息

Krupin T, Ritch R, Camras C B, Brucker A J, Muldoon T O, Serle J, Podos S M, Sinclair S H

机构信息

Department of Ophthalmology, Scheie Eye Institute, University of Pennsylvania School of Medicine, Philadelphia 19104.

出版信息

Ophthalmology. 1988 Sep;95(9):1174-80. doi: 10.1016/s0161-6420(88)33030-7.

Abstract

A long glaucoma valve implant attached to an external scleral explant was used during filtration surgery in 72 eyes: 39 eyes with neovascular glaucoma and 33 eyes with other types of secondary glaucomas or with primary glaucoma in which prior filtration surgery had failed. The implant consisted of an open Silastic tube (outside diameter, 0.64 mm), which was placed into the anterior chamber. The external end of the tube contained a pressure-sensitive (opening pressure, 11 mmHg) and unidirectional slit-valve, and was sutured within the groove of a #220 Silastic explant. The 180 degree explant was placed beneath three rectus muscles and then sutured so that the grooved side was against the sclera, with the anterior edge 8 to 12 mm posterior to the limbus. The long glaucoma valve implant resulted in a large, posterior bleb extending over the area of the Silastic explant. The mean preoperative intraocular pressure (IOP) of 43.9 mmHg in the eyes with neovascular glaucoma was reduced to 17.4 mmHg after a mean follow-up of 20.2 months. The mean preoperative IOP of 38.1 mmHg in the eyes after failure of previous filtration surgery was reduced to 17.6 mmHg at a mean follow-up of 21.0 months. Postoperative IOP was less than 21 mmHg in 77% of eyes with neovascular glaucoma (47% required additional medication) and in 82% of eyes with previous failure of filtration surgery (56% required additional medication).

摘要

在72只眼中,滤过手术时使用了连接外部巩膜外植体的长青光眼引流阀植入物:39只新生血管性青光眼眼,33只其他类型继发性青光眼眼或原发性青光眼且先前滤过手术失败的眼。植入物由一根开放的硅橡胶管(外径0.64毫米)组成,将其置于前房。管子的外端包含一个压力敏感型(开启压力11 mmHg)单向裂隙阀,并缝合在#220硅橡胶外植体的凹槽内。180度的外植体置于三条直肌下方,然后缝合,使带凹槽的一侧靠在巩膜上,前缘位于角膜缘后8至12毫米处。长青光眼引流阀植入物导致在硅橡胶外植体区域上方形成一个大的后部滤过泡。新生血管性青光眼眼术前平均眼压43.9 mmHg,平均随访20.2个月后降至17.4 mmHg。先前滤过手术失败的眼术前平均眼压38.1 mmHg,平均随访21.0个月后降至17.6 mmHg。新生血管性青光眼眼中77%的术后眼压低于21 mmHg(47%需要额外药物治疗),先前滤过手术失败的眼中82%的术后眼压低于21 mmHg(56%需要额外药物治疗)。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验