Anesthesiology and Intensive care, Università degli Studi di Sassari, Sassari, Italy.
Médecine Intensive Réanimation, C.H.U de Grenoble-Alpes, Grenoble, France.
PLoS One. 2020 Mar 11;15(3):e0230147. doi: 10.1371/journal.pone.0230147. eCollection 2020.
When patients with acute respiratory distress syndrome are moved out of an intensive care unit, the ventilator often requires changing. This procedure suppresses positive end expiratory pressure and promotes lung derecruitment. Clamping the endotracheal tube may prevent this from occurring. Whether or not such clamping maintains positive end-expiratory pressure has never been investigated. We designed a bench study to explore this further.
We used the Elysee 350 ventilator in 'volume controlled' mode with a positive end-expiratory pressure of 15 cmH2O, connected to an endotracheal tube with an 8 mm internal diameter inserted into a lung model with 40 ml/cmH2O compliance and 10 cmH2O/L/s resistance. We measured airway pressure and flow between the distal end of the endotracheal tube and the lung model. We tested a plastic, a metal, and an Extra Corporeal Membrane Oxygenation clamp, each with an oral/nasal, a nasal, and a reinforced endotracheal tube. We performed an end-expiratory hold then clamped the endotracheal tube and disconnected the ventilator. We measured the change in airway pressure and volume for 30 s following the disconnection of the ventilator.
Airway pressure decreased thirty seconds after disconnection with all combinations of clamp and endotracheal tube. The largest fall in airway pressure (-17.486 cmH2O/s at 5 s and -18.834 cmH2O/s at 30 s) was observed with the plastic clamp combined with the reinforced endotracheal tube. The smallest decrease in airway pressure (0 cmH2O/s at 5 s and -0.163 cmH2O/s at 30 s) was observed using the Extra Corporeal Membrane Oxygenation clamp with the nasal endotracheal tube.
Only the Extra Corporeal Membrane Oxygenation clamp was efficient. Even with an Extra Corporeal Membrane Oxygenation clamp, it is important to limit the duration the ventilator is disconnected to a few seconds (ideally 5 s).
急性呼吸窘迫综合征患者转出重症监护病房时,通常需要更换呼吸机。该操作会抑制呼气末正压并促进肺复张。气管内管夹闭可防止这种情况发生。气管内管夹闭是否能维持呼气末正压尚未得到研究。我们设计了一项基础研究来进一步探讨这个问题。
我们使用 Elysee 350 呼吸机,在呼气末正压为 15cmH2O 的情况下,采用容量控制模式,将内径为 8mm 的气管内管连接到一个顺应性为 40ml/cmH2O、阻力为 10cmH2O/L/s 的肺模型上。我们测量了气管内管远端与肺模型之间的气道压力和流量。我们测试了一种塑料夹、一种金属夹和一种体外膜肺氧合夹,每种夹都有一个口腔/鼻腔、一个鼻腔和一个加强型气管内管。我们进行了呼气末保持,然后夹闭气管内管并断开呼吸机。我们测量了断开呼吸机后 30 秒内气道压力和容量的变化。
所有气管内管和夹的组合在断开呼吸机后 30 秒内气道压力均下降。使用加强型气管内管的塑料夹时,气道压力下降最大(第 5 秒时为-17.486cmH2O/s,第 30 秒时为-18.834cmH2O/s)。使用鼻腔气管内管的体外膜肺氧合夹时,气道压力下降最小(第 5 秒时为 0cmH2O/s,第 30 秒时为-0.163cmH2O/s)。
只有体外膜肺氧合夹是有效的。即使使用体外膜肺氧合夹,也应将呼吸机断开的时间限制在几秒钟内(理想情况下为 5 秒)。