Comparison between predicted and actual dose distribution in interstitial therapy of the oral cavity.

作者信息

van 't Riet A, Veen R E, Mak A C, Stam P W, van Slooten F H, te Loo H J

机构信息

Radiotherapeutic Institute Stedendriehoek en Omstreken, Deventer, The Netherlands.

出版信息

Radiother Oncol. 1988 Dec;13(4):277-83. doi: 10.1016/0167-8140(88)90223-x.

Abstract

During the period 1985-1986, 10 patients with lesions of the floor of the mouth or of the mobile part of the tongue were treated by interstitial curietherapie with iridium-192. In the patients treated with the looping technique essential differences were found between the predicted source configuration and that actually achieved, resulting in a disagreement between precalculated and actual dose distribution. To avoid this undesirable situation, a simple oral cavity applicator was constructed which guarantees parallelism and equidistance between the implanted needles. It was found that the use of this applicator resulted in an almost complete agreement between forecast and actual dose distribution.

摘要

在1985年至1986年期间,对10例口底或舌活动部病变患者采用铱-192组织间近距离治疗。在用环形技术治疗的患者中,发现预测的源配置与实际实现的配置存在本质差异,导致预先计算的剂量分布与实际剂量分布不一致。为避免这种不理想的情况,制作了一种简单的口腔施源器,可确保植入针之间的平行度和等距性。结果发现,使用这种施源器可使预测剂量分布与实际剂量分布几乎完全一致。