Richter Manuel J, Harutyunova Satenik, Bollmann Tom, Classen Simon, Fuge Jan, Gall Henning, Gerhardt Felix, Ghofrani Hossein A, Gunkel Hartmut, Grünig Ekkehard, Halank Michael, Heine Alexander, Klose Hans, Lange Tobias J, Neurohr Claus, Nickolaus Kai, Opitz Christian F, Rosenkranz Stephan, Seyfarth Hans-Jürgen, Tello Khodr, Ewert Ralf, Olsson Karen M
Department of Pneumology, Kerckhoff Heart, Rheuma and Thoracic Center, Bad Nauheim, Germany.
Department of Internal Medicine, Justus Liebig University Giessen, Universities of Giessen and Marburg Lung Center (UGMLC), Giessen, Germany.
Pulm Circ. 2020 Mar 13;10(1):2045894020910136. doi: 10.1177/2045894020910136. eCollection 2020 Jan-Mar.
Implantable infusion pumps might improve the convenience and safety of intravenous treprostinil for pulmonary arterial hypertension. The LENUS Pro® pump (approved in Europe) has a fixed flow rate. Based on 126 pumps and 2853 refills, we retrospectively analyzed the actual flow rate from 09/2010 to 09/2018. A relevant flow rate variance is evident after three years; therefore, flow rate monitoring and dose adjustment are mandatory.