Radiation Safety and Quality Assurance Division, Hospital East, National Cancer Center, Kashiwa, Chiba, Japan.
School of Allied Health Sciences, Kitasato University, Kanagawa, Japan.
Brachytherapy. 2020 May-Jun;19(3):362-371. doi: 10.1016/j.brachy.2020.02.002. Epub 2020 Mar 21.
The purpose of this study was to develop a novel quality assurance (QA) program to check the entire treatment chain of image-guided brachytherapy with dose distribution evaluation in a single setup and irradiation using a gel dosimeter.
A polymer gel was used, and the readout was performed by magnetic resonance scanning. A CT-based treatment plan was generated using the Oncentra planning system (Elekta, Sweden), and irradiation was performed three times using an afterloading device with an Ir-192 source. The dose-response curve of the gel was created using 6-MV X-ray, which is independent of the source beams. Planar gamma images on a coronal plane along the source transport axis were calculated using the measured dose as a reference, and the calculated doses were used in several error simulations (no error; 2.0 or 2.5 mm systematic and random source dwell mispositioning; and dose error of 2%, 5%, 10%, and 20%).
The dose-R (spin-spin relaxation rate) conversion table revealed that the uncertainty and dose resolution of 6-MV X-ray were better than those of Ir-192 and also constant between the three measurements. With the 3%/1 mm criteria, there were statistically significant differences between each pair of settings except dose error of 2% and 5%.
This work depicts a simple and efficient end-to-end test that can provide a clinically useful tool for QA of image-guided brachytherapy. In this QA program, air kerma strength and dwell position setting could also be verified. This test can also distinguish between different types of error.
本研究旨在开发一种新的质量保证(QA)方案,以在单次设置中使用凝胶剂量计对图像引导近距离放射治疗的整个治疗链进行剂量分布评估,同时进行照射。
使用聚合物凝胶,通过磁共振扫描进行读出。使用 Oncentra 计划系统(Elekta,瑞典)生成基于 CT 的治疗计划,并使用带有 Ir-192 源的后装设备进行三次照射。凝胶的剂量响应曲线是使用 6-MV X 射线创建的,与源射线无关。使用测量剂量作为参考,计算沿源传输轴的冠状面上的平面伽马图像,并在几个误差模拟中使用计算剂量(无误差;2.0 或 2.5mm 系统和随机源驻留位置错误;以及 2%、5%、10%和 20%的剂量误差)。
剂量-R(自旋-自旋弛豫率)转换表显示,6-MV X 射线的不确定性和剂量分辨率优于 Ir-192,并且在三次测量之间也是恒定的。在 3%/1mm 标准下,除了 2%和 5%的剂量误差外,每个设置之间的差异均具有统计学意义。
这项工作描述了一种简单而有效的端到端测试,可以为图像引导近距离放射治疗的 QA 提供临床有用的工具。在这个 QA 方案中,空气比释动能和驻留位置设置也可以得到验证。该测试还可以区分不同类型的误差。