Rabaeh Khalid A, Al-Tarawneh Rawan E, Eyadeh Molham M, Hammoudeh Issra' M E, Shatnawi Moneeb T M
Medical Imaging Department, Faculty of Applied Medical Sciences, The Hashemite University, Zarqa 13133, Jordan.
Physics Department, School of Science, The University of Jordan, Amman 11942, Jordan.
Gels. 2022 Jan 26;8(2):78. doi: 10.3390/gels8020078.
The impact of calcium chloride (CaCl) on the performance of -(hydroxymethyl)acrylamide (NHMA) polymer gel dosimeter is studied in this article. The dosimeter was exposed to doses of up to 10 Gy with radiation beam-energy of 10 MV and dose-rates of 300 cGy/min. The relaxation rate (R) parameter was utilized to explore the performance of irradiated NHMAGAT gels. The dose response in terms of R increased from 0.29 to 0.63 Gy·s with increasing calcium chloride concentration from 0 to 1000 mM. The results show no substantial impact of dose-rates as well as radiation energies on NHMAGAT samples. For the steadiness of irradiated NHMAGAT dosimeters, it was found that there is no apparent variation in R (less than ±3%; standard deviation) up to 3 days. The overall uncertainty of the gel dosimeter with calcium chloride is 4.96% (double standard deviation, 95% confidence level).
本文研究了氯化钙(CaCl)对β-(羟甲基)丙烯酰胺(NHMA)聚合物凝胶剂量计性能的影响。该剂量计在10 MV的辐射束能量和300 cGy/min的剂量率下接受高达10 Gy的剂量照射。利用弛豫率(R)参数来探究辐照后的NHMAGAT凝胶的性能。随着氯化钙浓度从0增加到1000 mM,R的剂量响应从0.29增加到0.63 Gy·s。结果表明,剂量率和辐射能量对NHMAGAT样品没有实质性影响。对于辐照后的NHMAGAT剂量计的稳定性,发现直到3天R没有明显变化(小于±3%;标准偏差)。含氯化钙的凝胶剂量计的总体不确定度为4.96%(双标准偏差,95%置信水平)。