Suppr超能文献

验证一种用于介入放射学中皮肤剂量图计算的剂量跟踪软件。

Validation of a dose tracking software for skin dose map calculation in interventional radiology.

机构信息

Medical Physics Department, ASST Grande Ospedale Metropolitano Niguarda, Piazza Ospedale Maggiore 3, 20162 Milano, Italy.

Medical Physics Department, ASST Grande Ospedale Metropolitano Niguarda, Piazza Ospedale Maggiore 3, 20162 Milano, Italy.

出版信息

Phys Med. 2020 Apr;72:122-132. doi: 10.1016/j.ejmp.2020.03.008. Epub 2020 Apr 3.

Abstract

PURPOSE

Validate the skin dose software within the radiation dose index monitoring system NEXO[DOSE]® (Bracco Injeneering S.A., Lausanne, Switzerland). It provides the skin dose distribution in interventional radiology (IR) procedures.

METHODS

To determine the skin dose distribution and the Peak Skin Dose (PSD) in IR procedures, the software uses exposure and geometrical parameters taken from the radiation dose structured report and additional information specific to each angiographic system. To test the accuracy of the software, GafChromic® XR-RV3 films, wrapped under a cylindrical PMMA phantom, were irradiated with different setups. Calculations and films results are compared in terms of absolute dose and geometric accuracy, using two angiographic systems (Philips Integris Allura FD20, Siemens AXIOM-ArtisZeego).

RESULTS

Calculated and film measured PSD values agree with an average difference of 7% ± 5%. The discrepancies in dose evaluation increase up to 33% in lower dose regions, because the algorithm does not consider the out-of-field scatter contribution of the neighboring fields, which is more significant in these areas. Regarding the geometric accuracy, the differences between the simulated dose spatial distributions and the measured ones are<3 mm (4%) in simple tests and 5 mm (5%) in setups closer to clinical practice. Moreover, similar results are obtained for the two studied angiographic system vendors.

CONCLUSIONS

NEXO[DOSE]® provides an accurate skin dose distribution and PSD estimate. It will allow faster and more accurate monitoring of patient follow-up in the future.

摘要

目的

验证辐射剂量指数监测系统 NEXO[DOSE]®(瑞士洛桑 Bracco Injeneering S.A.)中的皮肤剂量软件。它提供介入放射学(IR)程序中的皮肤剂量分布。

方法

为了确定 IR 程序中的皮肤剂量分布和峰值皮肤剂量(PSD),该软件使用来自辐射剂量结构化报告的曝光和几何参数以及每个血管造影系统特有的其他信息。为了测试软件的准确性,将 GafChromic® XR-RV3 胶片包裹在圆柱形 PMMA 体模下,用不同的设置进行照射。使用两种血管造影系统(飞利浦 Integris Allura FD20、西门子 AXIOM-ArtisZeego),从绝对剂量和几何精度两个方面比较计算结果和胶片结果。

结果

计算得出的和胶片测量的 PSD 值平均差异为 7%±5%。在剂量较低的区域,评估差异增加到 33%,因为该算法没有考虑相邻射野的场外散射贡献,在这些区域更为显著。关于几何精度,模拟剂量空间分布与测量结果之间的差异在简单测试中<3mm(4%),在更接近临床实践的设置中为 5mm(5%)。此外,对于所研究的两种血管造影系统供应商,也得到了类似的结果。

结论

NEXO[DOSE]®可提供准确的皮肤剂量分布和 PSD 估计。它将允许更快、更准确地监测患者的后续治疗。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验