Suppr超能文献

环氧乙烷灭菌对聚乙烯醇水凝胶与γ射线辐照的影响。

Effect of Ethylene Oxide Sterilization on Polyvinyl Alcohol Hydrogel Compared with Gamma Radiation.

机构信息

Department of Chemical Engineering, University of Waterloo, Waterloo, Canada.

Department of Biomedical Engineering, Oregon Health & Science University, Portland, Oregon, USA.

出版信息

Tissue Eng Part A. 2020 Oct;26(19-20):1077-1090. doi: 10.1089/ten.TEA.2020.0002. Epub 2020 May 28.

Abstract

This study investigated the effects of terminal sterilization of polyvinyl alcohol (PVA) biomaterials using clinically translatable techniques, specifically ethylene oxide (EtO) and gamma (γ) irradiation. While a few studies have reported the possibility of sterilizing PVA with γ-radiation, the use of EtO sterilization of PVA requires additional study. PVA solutions were chemically crosslinked with trisodium trimetaphosphate and sodium hydroxide. The three experimental groups included untreated control, EtO, and γ-irradiation, which were tested for the degree of swelling and water content, and mechanical properties such as radial compliance, longitudinal tensile, minimum bend radius, burst pressure, and suture retention strength. In addition, samples were characterized with scanning electron microscopy, differential scanning calorimetry, X-ray photoelectron spectroscopy, and water contact angle measurements. Cell attachment was assessed using the endothelial cell line EA.hy926, and the sterilized PVA cytotoxicity was studied with a live/dead stain. Platelet and fibrin accumulation was measured using an shunt baboon model. Finally, the immune responses of PVA implants were analyzed after a 21-day subcutaneous implantation in rats and a 30-day implantation in baboon. EtO sterilization reduced the PVA graft wall thickness, its degree of swelling, and water content compared with both γ-irradiated and untreated PVA. Moreover, EtO sterilization significantly reduced the radial compliance and increased Young's modulus. EtO did not change PVA hydrophilicity, while γ-irradiation increased the water contact angle of the PVA. Consequently, endothelial cell attachment on the EtO-sterilized PVA showed similar results to the untreated PVA, while cell attachment significantly improved on the γ-irradiated PVA. When exposing the PVA grafts to circulating whole blood, fibrin accumulation of EtO-sterilized PVA was found to be significantly lower than γ-irradiated PVA. The immune responses of γ-irradiated PVA, EtO-treated PVA, and untreated PVA were compared. Implanted EtO-treated PVA showed the least MAC387 reaction. The terminal sterilization methods in this study changed PVA hydrogel properties; nevertheless, based on the characterizations performed, both sterilization methods were suitable for sterilizing PVA. We concluded that EtO can be used as an alternative method to sterilize PVA hydrogel material. Impact statement Polyvinyl alcohol (PVA) hydrogels have been used for a variety of tissue replacements, including neural, cardiac, meniscal, cartilage, muscle, pancreatic, and ocular applications. In addition, PVA can be made into a tubular shape and used as a small-diameter vascular graft. Ethylene oxide (EtO) is one of the Food and Drug Administration-approved methods for sterilization, but its effect on PVA has not been studied extensively. The outcome of this study provides the effects of EtO and γ-irradiation of PVA grafts on both the material properties and the responses, particularly for vascular applications. Knowledge of these effects may ultimately improve the success rate of PVA vascular grafts.

摘要

本研究探讨了使用临床上可转化的技术(特别是环氧乙烷(EtO)和γ(γ)辐照)对聚乙烯醇(PVA)生物材料进行终端灭菌的效果。虽然有一些研究报告了用γ辐射对 PVA 进行灭菌的可能性,但 EtO 对 PVA 的灭菌仍需要进一步研究。将 PVA 溶液用三偏磷酸钠和氢氧化钠化学交联。三个实验组包括未处理的对照、EtO 和γ辐照,对其膨胀度和含水量以及径向顺应性、纵向拉伸、最小弯曲半径、爆破压力和缝线保持强度等机械性能进行了测试。此外,还对样品进行了扫描电子显微镜、差示扫描量热法、X 射线光电子能谱和水接触角测量。使用内皮细胞系 EA.hy926 评估细胞附着,并用活/死染色研究经灭菌的 PVA 的细胞毒性。使用分流狒狒模型测量血小板和纤维蛋白的积累。最后,在大鼠皮下植入 21 天和狒狒植入 30 天后,分析 PVA 植入物的免疫反应。与γ辐照和未处理的 PVA 相比,EtO 灭菌降低了 PVA 移植物壁的厚度、膨胀度和含水量。此外,EtO 灭菌显著降低了径向顺应性并增加了杨氏模量。EtO 没有改变 PVA 的亲水性,而γ辐照增加了 PVA 的水接触角。因此,EtO 灭菌的内皮细胞附着在 PVA 上的结果与未处理的 PVA 相似,而γ辐照的 PVA 上的细胞附着显著改善。当将 PVA 移植物暴露于循环全血中时,发现 EtO 灭菌的 PVA 中的纤维蛋白积累明显低于γ辐照的 PVA。比较了γ辐照的 PVA、EtO 处理的 PVA 和未处理的 PVA 的免疫反应。植入的 EtO 处理的 PVA 显示出最低的 MAC387 反应。本研究中的终末灭菌方法改变了 PVA 水凝胶的性质;然而,根据所进行的特性分析,两种灭菌方法都适用于 PVA 的灭菌。我们得出结论,EtO 可以作为灭菌 PVA 水凝胶材料的替代方法。影响说明聚乙烯醇(PVA)水凝胶已用于多种组织替代物,包括神经、心脏、半月板、软骨、肌肉、胰腺和眼部应用。此外,PVA 可以制成管状形状并用作小直径血管移植物。环氧乙烷(EtO)是食品和药物管理局批准的灭菌方法之一,但尚未对其对 PVA 的影响进行广泛研究。这项研究的结果提供了 EtO 和γ辐照对 PVA 移植物的材料性能和反应的影响,特别是对血管应用的影响。了解这些影响最终可能会提高 PVA 血管移植物的成功率。

相似文献

5
Effects of sterilization on poly(ethylene glycol) hydrogels.灭菌对聚乙二醇水凝胶的影响。
J Biomed Mater Res A. 2008 Dec 1;87(3):608-17. doi: 10.1002/jbm.a.31811.

引用本文的文献

9
The effects of surface topography modification on hydrogel properties.表面形貌改性对水凝胶性能的影响。
APL Bioeng. 2021 Jul 27;5(3):031509. doi: 10.1063/5.0046076. eCollection 2021 Sep.
10
Regeneration of skeletal system with genipin crosslinked biomaterials.使用京尼平交联生物材料实现骨骼系统的再生
J Tissue Eng. 2020 Nov 29;11:2041731420974861. doi: 10.1177/2041731420974861. eCollection 2020 Jan-Dec.

本文引用的文献

5
Injectable hydrogels for cartilage and bone tissue engineering.用于软骨和骨组织工程的可注射水凝胶。
Bone Res. 2017 May 30;5:17014. doi: 10.1038/boneres.2017.14. eCollection 2017.