Niederle P, Stárek A, Jezek V, Jezková J, Hes I, Michaljanic A
Institute of Physiological Regulations, Czechoslovak Academy of Sciences, Prague.
Cor Vasa. 1988;30(5):338-44.
The authors used the pulse and continuous-wave Doppler technique to perform 51 measurements of stroke volume and cardiac output in 37 persons. Results were compared with those obtained using Fick's principle of thermodilution (immediately subsequent or simultaneous). Regression analysis showed a close correlation between Doppler and haemodynamic values - r = 0.86 for stroke volume and 0.94 for cardiac output (p less than 0.001). Using the paired t-test, the methods did not differ significantly. Standard deviation of Doppler stroke volume values from reference haemodynamic values was 10%, and only two measurements differed by more than 25%. Doppler determination of cardiac output can be therefore recommended as an alternative method in patients with high-quality echo Doppler recordings.
作者使用脉冲和连续波多普勒技术对37人进行了51次每搏量和心输出量测量。将结果与采用Fick热稀释原理(紧接着或同时)获得的结果进行比较。回归分析显示多普勒值与血流动力学值之间密切相关——每搏量的r值为0.86,心输出量的r值为0.94(p<0.001)。采用配对t检验,两种方法差异无统计学意义。多普勒每搏量值与参考血流动力学值的标准差为10%,只有两次测量差异超过25%。因此,对于有高质量回声多普勒记录的患者,推荐将多普勒心输出量测定作为一种替代方法。