Int J Environ Res Public Health. 2020 Apr 13;17(8):2651. doi: 10.3390/ijerph17082651.
Continuously evolving laser devices can be used in various fields; they are an alternative to the traditional cold blade surgery to perform biopsies of oral soft tissues. The aspect focused on in this paper is the possibility to use the 445 nm diode laser (Eltech K-Laser srl, Treviso, Italy) in complete safety, by evaluating its thermal effects during microscopy. A histological evaluation of the alteration of the peri-incisional edges on 10 samples was realized. All excisional biopsies were related to clinically unsuspected lesions and performed by the same expert operator. The surgical procedure was performed with the same laser parameters and the same pathologist evaluated the thermal effect on the samples. An average value of the detected tissue alteration was calculated; the average damage of the epithelium was 650.93 μm, while in the connective tissue it was 468.07 μm. In all the cases a clear diagnosis was possible, and no clinical complications were observed; so, the 445 nm diode laser proved to be a device that can be safely used for biopsies of clinically unsuspicious lesions. Due to the small number of cases, this in vivo preliminary experience needs to be extended.
不断发展的激光设备可应用于多个领域;它们是传统冷刀手术的替代品,可用于对口腔软组织进行活检。本文关注的是使用 445nm 半导体激光(Eltech K-Laser srl,特雷维索,意大利)的可能性,通过在显微镜下评估其热效应,可以完全安全地使用该激光。对 10 个样本的切口边缘进行了组织学评估。所有切除活检均与临床无明显异常的病变相关,由同一位专家操作员完成。手术过程采用相同的激光参数进行,同一位病理学家评估样本的热效应。计算出检测到的组织改变的平均值;上皮的平均损伤为 650.93μm,而在结缔组织中为 468.07μm。在所有情况下均可以明确诊断,且未观察到临床并发症;因此,445nm 半导体激光被证明是一种可安全用于临床无明显异常病变活检的设备。由于病例数量较少,这种体内初步经验需要进一步扩展。