Dobbie Laurence J, Mackin Sharon T, Hogarth Katrina, Lonergan Frances, Kannenkeril Dennis, Brooksbank Katriona, Delles Christian
Institute of Cardiovascular and Medical Sciences, University of Glasgow, Glasgow, UK.
Department of Nephrology and Hypertension, University of Erlangen-Nürnberg, Germany.
Blood Press Monit. 2020 Aug;25(4):216-223. doi: 10.1097/MBP.0000000000000448.
Flow-mediated dilation (FMD) is a non-invasive imaging modality used to measure endothelial function but has significant intra- and inter-observer variability. The use of semi-automated FMD devices could overcome this limitation. We assessed the reproducibility of same-day semi-automated FMD measurements by investigators who received basic training on the correct use of the device.
Forty-three healthy volunteers had two brachial artery FMD measurements performed 20 minutes apart using the UNEX EF 38G device, and automated outputs were produced. Images were also manually analysed using edge-detection software. The reproducibility of repeat FMD measurements within individuals was compared for automated and manual readings, and the correlation between analytical techniques was calculated.
Twenty-five percent of scans were of non-diagnostic quality (n = 32). Automated analyses demonstrated sub-optimal reproducibility and measurement variability [intraclass correlation coefficient (ICCC) = 0.334, coefficient of variation (CV) = 45.87%]. In contrast, manually analysed scans had excellent reproducibility and low measurement variance (ICCC = 0.815, CV = 11.40%). FMD values obtained from automated and manual analysis correlated poorly (r = 0.164), whereas resting (r = 0.955) and maximal brachial artery diameters demonstrated excellent correlation (r = 0.867).
Manually evaluated serial UNEX EF readings have good reproducibility and therefore, the optimal FMD workflow involves manual analyses prior to independent automated interrogation. The high non-diagnostic scan rate is most likely the result of insufficient training and indicates that semi-automatic devices such as UNEX EF should be used by experienced investigators to achieve optimal results.
血流介导的血管舒张功能(FMD)是一种用于测量内皮功能的非侵入性成像方法,但在观察者内和观察者间存在显著差异。使用半自动FMD设备可以克服这一局限性。我们评估了接受过该设备正确使用基础培训的研究人员在同一天进行的半自动FMD测量的可重复性。
43名健康志愿者使用UNEX EF 38G设备,间隔20分钟进行两次肱动脉FMD测量,并生成自动输出结果。还使用边缘检测软件对图像进行手动分析。比较个体内重复FMD测量在自动和手动读数方面的可重复性,并计算分析技术之间的相关性。
25%的扫描质量无法用于诊断(n = 32)。自动分析显示可重复性和测量变异性欠佳[组内相关系数(ICCC)= 0.334,变异系数(CV)= 45.87%]。相比之下,手动分析的扫描具有出色的可重复性和低测量方差(ICCC = 0.815,CV = 11.40%)。自动和手动分析获得的FMD值相关性较差(r = 0.164),而静息状态(r = 0.955)和肱动脉最大直径显示出极好的相关性(r = 0.867)。
手动评估的连续UNEX EF读数具有良好的可重复性,因此,最佳的FMD工作流程包括在独立自动检测之前进行手动分析。高非诊断性扫描率很可能是培训不足的结果,这表明经验丰富的研究人员使用如UNEX EF这样的半自动设备才能获得最佳结果。