新型压平眼压计用于激光屈光手术后的近视患者。

New applanation tonometer for myopic patients after laser refractive surgery.

机构信息

Instituto Universitario Barraquer, Barraquer Ophthalmology Centre, Barcelona, 08012, Spain.

Department of Physics, University of Vigo, 36310, Galicia, Spain.

出版信息

Sci Rep. 2020 Apr 27;10(1):7053. doi: 10.1038/s41598-020-64013-4.

Abstract

This study assesses the agreement between intraocular pressure (IOP) measurements taken with the Goldmann applanation tonometer (GAT) and a new experimental applanation tonometer with a convexly shaped apex (CT) after laser myopic refractive surgery. Two different CT radii (CT1 and CT2) were designed with a finite element analyser, and a prospective double masked study on 102 eyes from 102 patients was carried out. A Bland-Altman plot and intra-class correlation coefficient (ICC) were calculated to assess the agreement between GAT measurements and the measurements of both CT1 and CT2 before and after myopic laser assisted in situ keratomileusis (LASIK; n = 73) and photorefractive keratectomy (PRK; n = 29). We evaluated a subset of two subgroups (n = 36 each) for intra and inter-observer (IA/IE) error. From the whole cohort, the best IOP agreement was observed between GATpre and CT1post surgery: 16.09 ± 2.92 vs 16.42 ± 2.87 (p < 0.001); ICC = 0.675 (95% CI: 0.554-0.768). In the analysis of LASIK vs PRK, GATpre and CT1post showed the highest agreement, although LASIK measurements were more accurate than PRK, as the ICC = 0.718 (95% CI: 0.594-0.812) and ICC = 0.578 (95% CI: 0.182-0.795) respectively. Excellent agreement was observed for IA/IE, and there was an ICC > 0.8 (95% CI) in all cases. CT1 proved more accurate in the LASIK subgroup. In conclusion, our new version of GAT could be used with post-surgery LASIK patients as a more accurate measurement device compared to the current reference tonometer.

摘要

这项研究评估了在激光近视屈光手术后,使用 Goldmann 压平眼压计 (GAT) 和一种新的具有凸形顶点的实验性压平眼压计 (CT) 测量的眼压 (IOP) 之间的一致性。两种不同的 CT 半径 (CT1 和 CT2) 使用有限元分析器设计,并对 102 名患者的 102 只眼进行了前瞻性、双盲研究。计算 Bland-Altman 图和组内相关系数 (ICC) 来评估 GAT 测量值与 CT1 和 CT2 测量值之间的一致性,这些测量值在激光辅助原位角膜磨镶术 (LASIK; n = 73) 和光折射性角膜切除术 (PRK; n = 29) 前后。我们评估了两个亚组(每组 n = 36)的观察者内和观察者间误差 (IA/IE)。在整个队列中,GATpre 和 CT1post 手术后的眼压一致性最好:16.09 ± 2.92 vs 16.42 ± 2.87 (p < 0.001);ICC = 0.675 (95%CI: 0.554-0.768)。在 LASIK 与 PRK 的分析中,GATpre 和 CT1post 显示出最高的一致性,尽管 LASIK 测量更准确,因为 ICC = 0.718 (95%CI: 0.594-0.812) 和 ICC = 0.578 (95%CI: 0.182-0.795)。IA/IE 观察到极好的一致性,并且在所有情况下,ICC > 0.8 (95%CI)。CT1 在 LASIK 亚组中更准确。总之,与当前的参考眼压计相比,我们的 GAT 新版本可用于 LASIK 术后患者,作为更准确的测量设备。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/7892/7184562/db96a4b7ccca/41598_2020_64013_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验