不同类型屈光手术后帕斯卡动态轮廓眼压计与压平眼压计的比较。

Comparison between Pascal dynamic contour tonometer and Goldmann applanation tonometer after different types of refractive surgery.

机构信息

Ophthalmology Department, Democritus University of Thrace, Alexandroupolis, Greece.

出版信息

Graefes Arch Clin Exp Ophthalmol. 2011 May;249(5):767-73. doi: 10.1007/s00417-010-1431-9. Epub 2010 Jun 25.

Abstract

BACKGROUND

To evaluate and compare the recorded IOP values of the Pascal dynamic contour tonometer (PDCT) and the Goldmann applanation tonometer (GAT) after photorefractive keratectomy (PRK) and laser in situ keratomileusis (LASIK).

METHODS

Three groups of 84, 182 and 43 patients each were treated by PRK for myopia, LASIK for myopia and LASIK for hyperopia respectively. Intraocular pressure (IOP) measurements were performed in all eyes 1 day before and 1, 3, 6 and 12 months after treatment with PDCT and GAT. Ultrasound pachymetry was performed in all eyes preoperatively and at 1st month postoperatively.

RESULTS

Preoperatively and postoperatively, GAT readings were lower than PDCT in all groups (all p < 0.05). Postoperatively in the PRK group, compared to the preoperative value, the mean differences of IOP recorded with GAT at 1st, 3rd, 6th and 12th month were -1.4 mmHg, -1.7 mmHg, -1.7 mmHg and -1.9 mmHg respectively (all p < 0.05). In the myopic LASIK group; the corresponding values with GAT were -3.6 mmHg, -3.6 mmHg, -3.6 mmHg and -3.5 mmHg (all p < 0.05), while in the hyperopic LASIK group the corresponding values were -1.1 mmHg, -0.7 mmHg, -1.1 mmHg and -0.9 mmHg (all p < 0.05). The mean IOP difference (GAT-PDCT) for myopic PRK, myopic LASIK and hyperopic LASIK were respectively -3.8 mmHg, -4.1 mmHg and -1.5 mmHg at the 12-month follow-up. No statistically significant changes were found for any group with the PDCT.

CONCLUSIONS

GAT-determined IOP values were significant lower at all time-points after hyperopic LASIK, as well as myopic PRK or LASIK. The Pascal tonometry values remained unaffected for all groups.

摘要

背景

评估和比较经光折射性角膜切削术(PRK)和准分子激光原位角膜磨镶术(LASIK)后帕斯卡动态轮廓眼压计(PDCT)和压平眼压计(GAT)记录的眼压值。

方法

分别对 84、182 和 43 例近视行 PRK、近视行 LASIK 和远视行 LASIK 的患者进行三组治疗。所有患者在治疗前 1 天及治疗后 1、3、6 和 12 个月用 PDCT 和 GAT 测量眼压。所有患者术前及术后 1 个月行超声角膜测厚术。

结果

在所有组中,术前和术后 GAT 读数均低于 PDCT(均 p<0.05)。在 PRK 组中,与术前相比,GAT 在术后 1、3、6 和 12 个月的平均眼压差值分别为-1.4mmHg、-1.7mmHg、-1.7mmHg 和-1.9mmHg(均 p<0.05)。在近视 LASIK 组中,GAT 的相应值分别为-3.6mmHg、-3.6mmHg、-3.6mmHg 和-3.5mmHg(均 p<0.05),而在远视 LASIK 组中,相应值分别为-1.1mmHg、-0.7mmHg、-1.1mmHg 和-0.9mmHg(均 p<0.05)。在 12 个月随访时,近视 PRK、近视 LASIK 和远视 LASIK 的平均眼压差值(GAT-PDCT)分别为-3.8mmHg、-4.1mmHg 和-1.5mmHg。PDCT 组未发现任何有统计学意义的变化。

结论

在远视 LASIK 以及近视 PRK 或 LASIK 后,所有时间点 GAT 测量的眼压值均显著降低。所有组的帕斯卡眼压计值均未受影响。