新的 Goldmann 凸面棱镜测量近视激光屈光手术后的眼压:与 GAT 和 ORA 的相关性。

Intraocular pressure after myopic laser refractive surgery measured with a new Goldmann convex prism: correlations with GAT and ORA.

机构信息

Instituto Universitario Barraquer, Barraquer Ophthalmology Centre, Laforja 88, 08012, Barcelona, Spain.

Department of Ophthalmology, Hospital Universitario Puerta De Hierro, 28222, Madrid, Spain.

出版信息

BMC Ophthalmol. 2022 Feb 16;22(1):79. doi: 10.1186/s12886-022-02309-x.

Abstract

BACKGROUND

The purpose of this study is to describe measurements using a newly developed modified Goldmann convex tonometer (CT) 1 year after myopic laser refractive surgery. Intraocular pressure (IOP) measurements were compared with IOP values obtained by Goldmann applanation tonometer (GAT), and Ocular Response Analyzer (ORA).

METHODS

Prospective double-masked study performed on thirty eyes of thirty patients that underwent laser in situ keratomileusis (LASIK; n = 19) or photorefractive keratectomy (PRK; n = 11). IOP was measured before and 3 and 12 months after surgery. Intraclass correlation coefficient (ICC) and Bland-Altman plot were calculated to assess the agreement between GAT, CT, IOPg (Goldmann-correlated IOP) and IOPcc (corneal-compensated IOP) from ORA.

RESULTS

Twelve months after LASIK, IOP measured with CT showed the best correlation with IOP measured with GAT before surgery (GATpre) (ICC = 0.886, 95% CI: 0.703-0.956) (15.60 ± 3.27 vs 15.80 ± 3.22; p < 0.000). However, a moderate correlation was found for IOP measured with IOPcc and CT 12 months after LASIK (ICC = 0.568, 95% CI: - 0.185 - 0.843) (15.80 ± 3.22 vs 12.87 ± 2.77; p < 0.004). Twelve months after PRK, CT showed a weak correlation (ICC = - 0.266, 95% CI: - 3.896 - 0.663), compared to GATpre (17.30 ± 3.47 vs 16.01 ± 1.45; p < 0.642), as well as poor correlation (ICC = 0.256, 95% CI: - 0.332 - 0.719) with IOPcc (17.30 ± 3.47 vs 13.38 ± 1.65; p < 0.182).

CONCLUSIONS

Twelve months after LASIK, IOP measured with CT strongly correlated with GAT before surgery and could therefore provide an alternative method for measuring IOP after this surgery. More studies regarding this new convex prism are needed to assess its accuracy.

摘要

背景

本研究旨在描述近视激光屈光手术后 1 年使用新开发的改良 Goldmann 凸面眼压计(CT)进行的测量。比较了眼压(IOP)测量值与 Goldmann 压平眼压计(GAT)和眼反应分析仪(ORA)获得的 IOP 值。

方法

对 30 例 30 只眼进行前瞻性双盲研究,这些患者接受了激光原位角膜磨镶术(LASIK;n=19)或光性角膜切削术(PRK;n=11)。在术前、术后 3 个月和 12 个月测量 IOP。计算了组内相关系数(ICC)和 Bland-Altman 图,以评估 GAT、CT、IOPg(Goldmann 相关眼压)和 ORA 中的 IOPcc(角膜补偿眼压)之间的一致性。

结果

LASIK 术后 12 个月,CT 测量的 IOP 与术前 GAT 测量的 IOP 相关性最好(GATpre)(ICC=0.886,95%CI:0.703-0.956)(15.60±3.27 与 15.80±3.22;p<0.000)。然而,LASIK 术后 12 个月,CT 测量的 IOP 与 IOPcc 的相关性为中度(ICC=0.568,95%CI:-0.185-0.843)(15.80±3.22 与 12.87±2.77;p<0.004)。PRK 术后 12 个月,CT 与 GATpre 相比呈弱相关性(ICC=-0.266,95%CI:-3.896-0.663)(17.30±3.47 与 16.01±1.45;p<0.642),与 IOPcc 的相关性也较差(ICC=0.256,95%CI:-0.332-0.719)(17.30±3.47 与 13.38±1.65;p<0.182)。

结论

LASIK 术后 12 个月,CT 测量的 IOP 与术前 GAT 高度相关,因此可为术后 IOP 测量提供替代方法。需要更多关于这种新型凸面棱镜的研究来评估其准确性。