Lee Ki-Young, Shin Kyung Su, Jung Ji-Hye, Cho Hye-Won, Kwon Kyung-Hwan, Kim Yu-Lee
Departments of Prosthodontics, Wonkwang University Dental Hospital, College of Dentistry, Wonkwang University, Iksan, Korea.
Departments of Oral and Maxillofacial Surgery, Wonkwang University Dental Hospital, College of Dentistry, Wonkwang University, Iksan, Korea.
J Korean Assoc Oral Maxillofac Surg. 2020 Apr 30;46(2):133-142. doi: 10.5125/jkaoms.2020.46.2.133.
In this study, we determined the incidence and pattern of screw loosening in patients who received dental implants.
Patients who received implants between January 2008 and October 2013 and completed their prosthetic rehabilitation were evaluated for the incidence, frequency, and onset of screw loosening using dental charts and radiographs. The association between each factor and screw loosening was analyzed using the chi-square test and a multivariate analysis with binary logistic regression models (<0.05).
Total 1,928 implants were placed in 837 patients (448 males, 389 females), whose follow-up period after loading varied from 0.25 to 70 months (mean period, 31.5 months). Screw loosening occurred in 7.2% of implants. Most cases occurred less than six months after loading. Among those, 22.3% experienced recurrent screw loosening. Screw loosening was most common in the molar region (8.5%) and frequently associated with an implant diameter of ≥5 mm (14.2%). External implant-abutment connections (8.9%) and screw-retained implant prostheses (10.1%) showed higher incidence of problems than internal implant-abutment connections and cement-retained implants, respectively. Screw loosening was most common in implant prostheses with single crowns (14.0%).
Within the limits of the current study, we conclude that the incidence of screw loosening differs significantly according to the position of implant placement, the type of implant and manufacturer, implant diameter, the type of implant-abutment connection, the type of retention in the implant prosthesis, and the type of implant prosthesis.
在本研究中,我们确定了接受牙种植体患者中螺钉松动的发生率和模式。
对2008年1月至2013年10月期间接受种植体植入并完成修复的患者,使用牙科图表和X光片评估螺钉松动的发生率、频率和发生时间。使用卡方检验和二元逻辑回归模型的多变量分析(<0.05)分析每个因素与螺钉松动之间的关联。
共对837例患者(448例男性,389例女性)植入了1928颗种植体,其加载后的随访期为0.25至70个月(平均31.5个月)。7.2%的种植体发生了螺钉松动。大多数病例发生在加载后不到六个月。其中,22.3%经历了复发性螺钉松动。螺钉松动在磨牙区最为常见(8.5%),且常与种植体直径≥5mm相关(14.2%)。种植体-基台外部连接(8.9%)和螺钉固位种植体修复体(10.1%)的问题发生率分别高于种植体-基台内部连接和粘结固位种植体。螺钉松动在单冠种植体修复体中最为常见(14.0%)。
在本研究的范围内,我们得出结论,螺钉松动的发生率根据种植体植入位置、种植体类型和制造商、种植体直径、种植体-基台连接类型、种植体修复体的固位类型以及种植体修复体类型的不同而有显著差异。